This is only a preview of the June 1989 issue of Silicon Chip. You can view 44 of the 96 pages in the full issue, including the advertisments. For full access, purchase the issue for $10.00 or subscribe for access to the latest issues. Articles in this series:
Articles in this series:
Articles in this series:
Articles in this series:
|
To effectively use an
oscilloscope, you must
know how the scope's
probe affects the signal
waveform you are
measuring. Here's a
rundown on how scope
probes work.
By JONATHAN GORDON
UNDERSTANDING
OSCILLOSCOPE PROBES
The most common type of
oscilloscope probe used by technicians is the voltage-sensing passive
probe. Because passive probes are
so common and used so routinely,
their peculiar effect on electronic
circuits is universally experienced
- though very often misunderstood.
Nothing can replace the kind of
troubleshooting knowledge that
comes from viewing different
waveforms from various types of
circuits, such as digital, analog and
radio frequency (RF). However,
understanding how an oscilloscope
probe loads the circuit under ·test
will give you a new troubleshooting
skill that no amount of experience
can equal.
Fig.1 is an equivalent circuit of a
typical xlO scope probe and the
vertical input of the associated
oscilloscope. The probe's head has
a 9-megohm resistor (Rl) that is
shunted by a 4-Z0pF trimmer
capacitor, Cl. The shielded coaxial
cable that connects the probe to the
scope has a distributed capacitance, C3, of approximately 80pF.
An electronic signal travels
through the probe tip, the probe
head and compensating network,
the shielded coaxial cable, and then
to the scope's input connector. The
input impedance of a typical scope
is 1 megohm (RZ) shunted by a Z0pF
capacitance (CZ). The scope's input
characteristics are often printed
near the input connector.
The idea of compensating a
passive probe is to balance the pro-
COMPENSATION
C1
CAO
PROBE
4-2OpF
PROBE
TIP
\
PROBE
HEAD
__ l_
---1
+
COAXIAL
CABLE
___\___
I
I
CAO
INPUT
I
r--~~I-----q-
- - - - - - r'\
-:-
I
C2
20pf1
R2
1M
.,.
FIG.1: INPUT EQUIVALENT CIRCUIT of a x10 probe and oscilloscope. The
probe's head includes a 9MO resistor which is shunted by a 4-20pF trimmer
capacitor {Cl) to provide compensation.
4
SILICON CHIP
be impedance and the scope impedance so their time constants are
equal, as shown in the following
formula:
RlCl = RZ(CZ + C3)
If their time constants are equal,
electrical waveforms will be communicated from the probe tip to the
scope input without the probe adding distortion to the signal. The
amplitude of the displayed pulse
will merely depend on the rnsistance ratio:
Vscope = (RZ/(Rl + RZ))Vinput
For example, using a xlO passive
probe, a lV p-p input at the probe's
tip will yield a 0.1 V p-p reading on
the scope's display. The resulting
decade attenuation of the xl0 probe is highly desirable because it affords a greater tip resistance (10
times the 1-megohm scope input
resistance) to minimise circuit
loading.
Once the probe has been compensated by adjusting trimmer capacitor Cl, the probe and scope input
are further reduced to the
equivalent circuit shown in Fig.2.
Copyright 1989, Gernsback Publications. Reprinted with permission from
January 1 989 Radio-Electronics.
PROBE
TIP
\
Another compensation scheme is
shown in Fig.5 which uses a 4-20pF
timmer capacitor to shunt the
resistor in the probe's head.
Pulse waveforms
Fig.2: INPUT EQUIVALENT CIRCUIT
of a xl0 compensated probe as seen
at the probe's tip.
Any circuit under measurement
will now see a single impedance at
the probe's tip of 10 megohms, Rp,
shunted by an 11.2pF capacitor, Cp,
which is a specification that is often
printed on the probe's head or compensation box.
Scope probe manufacturers have
been clever in the methods they use
to shunt the resistor in the probe's
body. For example, Fig.3 shows the
inside of a Tektronix Model 6006
xlO probe. The coaxial cable's centre conductor is extended into the
probe's body where it connects to a
cylinder that slides over the
resistor. At one end of the resistor
is a shorting slug that makes contact with the cylinder. A capacitor
is thus formed by the cylinder and
the resistor.
The probe is compensated by
screwing the probe's body into the
locking sleeve , which is then
tightened. During that adjustment,
the inner cylinder slides a varying
distance over the resistor and this
varies the shunt capacitance. Note
that the coaxial cable 's outer conductor is connected to a second
cylinder for shielding.
Fig.4 shows the inside of a
Tektronix Model P6105A xl0 probe. In this unit the cylindrical tubing forms a fixed shunt capacitance
across the resistor. No probe adjustment can be made in the probe's
body to compensate the probe
assembly. Instead the assembly
uses a compensation box at the connector end that houses a trimmer
capacitor, C2, which is connected
from the centre conductor to
ground. The shunt capacitance C3
and trimmer C2 are in parallel.
A common variation of this compensation arrangement uses a trimmer in the BNC plug housing for the
probe. This is used by probe
manufacturers such as Coline and
Hitachi.
In general, probes are compared
by how well they transfer an electrical pulse to the oscilloscope's input without causing distortion of
one kind or another. But what is the
real nature of an electrical pulse?
Fig.6(a) shows what a rec-
tangular pulse looks like on an
oscilloscope as amplitude vs. time.
However, as shown in Fig.6(b), the
same rectangular pulse displayed
on a spectrum analyser is transformed into a chart of amplitude vs.
frequency - ie, it shows what the
pulse would look like if broken up
into its individual sinusoidal harmonic components. As you can see,
the rectangular pulse is made up of
both even and odd order harmonic
components.
CYLINDRICAL
SLEEVEE
LOCKING
SLEEVE
ALLIGATOR
GROUNDING CLIP
Fig.3: THIS PROBE IS COMPENSATED by adjusting the locking sleeve.
C3 COAXIAL CABLE
SHUNT CAPACITANCE
·,
INNE,,
CONDUCTOR
CYLINDRICAL SLEEVE FORMING
FIXED SHUNT CAPACITOR
ACROSS RESISTOR
,::;,:._/
ALLIGATOR GROUNDING CLIP
FIG.4: THIS PROBE IS COMPENSATED by adjusting a trimmer capacitor
housed in the compensation box at the end of the probe's cable.
JUNE 1989
5
Rl
COAXIAL SHIELDED
CABLE
9M
,
I
,
I
I
I
\
t
\
=...._
PPOBE
TIP
\
TRIMMERIADJUSTMENT
SCREW
FIG.5: THIS PROBE IS COMPENSATED by adjusting a trimmer capacitor
housed in the probe's head.
w
Cl
:::,
PULSE
w
PULSE
Q
:::,
...:i
t:
~
"'"'
"'"'
......
TIME
(a}
:,:;
:,:;
"'
en
FREQUENCY
(bJ
FIG.6: THE PULSE WAVEFORM in (a) shows how an oscilloscope displays an
. electrical pulse as amplitude vs. time. As shown in (b ), the same pulse
waveform is displayed on spectrum analyser as amplitude vs. frequency.
Unlike the rectangular pulse, a
perfect square wave is made up of
only odd-order harmonics (that is 1,
3, 5, 7, 9, etc). For example, if a
lkHz square wave is input to an
oscilloscope, the lkHz fundamental
(1st harmonic) sinusoid up to the
9kHz harmonic sinusoid must be
reproduced without attenuation or
phase shifting.
As shown in Fig.7(a), the 1st' and
3rd harmonic components produce
a rather poor square wave. In
Fig.7(b), the square wave looks a little better when the 1st, 3rd and 5th
harmonics are present. And, as
shown in Fig.7(c), the square wave
looks better still when the 1st to 7th
odd-order harmonic components
are present. The waveform will appear sufficiently square and undistorted when frequencies are present out to about the 9th harmonic
sinusoid.
Obviously, the shape of a square
wave displayed on an oscilloscope
depends on the amplitude and
phase relationships of the harmonic
components. To accurately reproduce an electrical pulse it would be
6
SILICON CHIP
necessary to design a circuit that
responds equally well to an infinite
number of harmonic frequencies ,
so that all harmonics are included.
In practice, that cannot be done so
a compromise between pulse shape
and circuit design must be made.
Compensation
Every electronics technician has
recorded pulse waveform data such
as risetime, width, amplitude and
repetition rate, only to realise too late - that the probe wasn't
compensated. Because an improperly compensated probe can
distort an otherwise perfect
waveform, the probe's compensation capacitor must be correctly adjusted and the test data remeasured.
Let's now examine how to properly compensate a probe and, additionally, how faulty compensation
can affect the measurement of
pulse waveforms.
Because probes should often be
compensated, most scopes provide
a square-wave calibration signal
accessible from the front panel.
There are other types of probe
calibrators, such as a linefrequency calibrator, a lkHz
square-wave calibrator and other
more exotic types. However, they
are used less often than the more
convenient front-panel scope
calibrators.
The front-panel probe calibration
signal is a lkHz repetitive square
wave. A lkHz square wave is used
because the sinusoidal harmonic
components are very close together, so the slightest offset in the
probe's frequency response will affect the amplitude and phase relationship of many harmonic components at once, resulting in a
visually distorted waveform.
Fig.8(a) shows a lkHz calibration
signal from a properly compensated probe. Notice that the square
wave is undistorted. Fig.8(b) shows
the same calibration signal from an
over-compensated probe. Here, the
leading edge high-frequency harmonics are passed while some of
the lower-frequency harmonics are
slightly attenuated. Some phase
shifting has also occurred. The
greater the drooping effect of the
waveform, the greater the degree
of low-frequency harmonic attenuation and phase shifting that has
occurred.
Fig.8(c) shows the same calibration signal from an under compensated probe. Notice the lack of high
frequency components in the
leading edge of the square wave (indicated by rounding). Now let's examine the relationship between
pulse shape, rise time and the
capacitance of the circuit.
When the driving pulse has a
slow rise time, or the pulse width is
comparatively wide, the stray
capacitance can have a fairly high
value without producing visible
distortion. That's because slower
rise times and greater pulse widths
correspond to fewer high frequency harmonics.
However, the same value of stray
capacitance can become intolerable when the driving pulse
has extremely fast rise times or a
very narrow width. The waveshape
then depends critically on the
preservation of high-frequency harmonics. As more and more stray
shunt capacitance is added to the
circuit under test, the shunt
(bypass) capacitive reactance
decreases in value. Harmonic frequencies that comprise the pulse's ·
edge will now be shorted to ground
by the lower shunt-capacitive
reactance.
Assume, for example, that a computer circuit is working just fine
and that you want to observe the
20MHz master clock. So you connect your xl scope probe - and the
whole system crashes. The clock's
waveform displayed on the scope
looks a little distorted. You then
remove the probe from the circuit
and the system immediately comes
up. What happened?
FUNDAMENTAL PLUS
3RD HARMONIC
'
sistive attenuation ratio, the lower the probe-tip shunt capacit-ance.
For example, xl, xlO and xl00 probe attenuation factors might have a
54pF, 11.2pF and 2pF tip capactance values, respectively.
Continuous wave
(a)
FUNDAMENTAL
....I
(b)
(a)
FUNDAMENTAL PLUS 3RD
AND 5TH HARMONICS
✓-\
\
\
\
\
\
(b)
(c)
FIG.8: A PROPERLY COMPENSATED
FUNOAMENT AL PLUS 3RD,
5TH AND 7TH HARMONICS
I
I
I
\
\
(c)
FIG.7: ALL PULSE WAVEFORMS are
built up from sinusoidal harmonic
components. As shown in (a), the
fundamental (1st harmonic) plus the
3rd harmonic create a rather
distorted square wave. In (b), the 5th
harmonic is added, thus creating a
less distorted square wave. In (c), the
7th harmonic is added, thus
minimising distortion even further.
probe will display a lkHz calibration
signal as a perfect square wave as
shown in (a). However, in (b), when
the probe is over compensated, the
same lkHz signal shows lowfrequency attenuation and phase
shifting as indicated by the drooping
effect. As shown in (c), when the
probe is under compensated, the
high-frequency components are lost,
as indicated by the rounded leading
edge.
One possible explanation is that
the added shunt capacitance of the
probe degraded the clock's rise
time, which threw off the system
timing. As shown in Fig.9, slower
rise time translates into a wider
pulse width. In this situation, you
should try using a xlO probe
instead of a xl probe because the
xlO probe has a lower shunt capacitance.
In general, the greater the re-
When measuring a continuous
signal from the output of a
sinewave oscillator, the probe-tip's
capacitive reactance (X_p) at the
operating frequency should be
considered.
In Fig.10, the total probe tip impedance, designated Zp, includes
the probe's resistive, capacitive
and inductive elements. The
capacitive and resistive elements
make up most of the probe's impedance. However, some probes
also include additional inductive
elements that are designed into the
probe itself to offset the capacitive
loading. For worst case analysis,
use the probe's capacitive reactance formula:
Xp = 1/2-n-fC
where C is the probe-tip capacitance. This value is often marked
somewhere on the body of the
probe.
For example, the Tektronix
model P6105A passive probe has a
10-megohm input resistance with a
tip capacitance of 11.2pF. The Xp
will equal 290 ohms at 50MHz.
Depending on the impedance of the
source, the probe's loading could
have a major effect on the signal
amplitude and possibly interfere
with the operation of the circuit
under test.
The typical curves for probe impedance vs. frequency vary for each
probe type so consult your
probe's specifications. For sinewave amplitude measurements, a
probe should have the highest
possible impedance at the frequency of interest.
Voltage derating
The maximum voltage (DC and
AC) that can be safely handled by a
probe varies with frequency. Fig.11
shows the voltage derating curve
for a Tektronix Model P6105A
passive probe. The curve may be
summarised by saying that the maximum voltage handling capability is
JUNE 1989
7
AMPLITUDE
10M
RISE TIME
It=
I
90 %
X'
50 %
I
10 ¾
\
I
/
1M
\
I
.;;-
\
;l
I\
e,
I
I~
-1
''
100k
'
''
=
><
~
<
\
10k
~
WIDTH
WIDTH
~
·,
\zp
\
'Xp' ' ,
'
1k
''
I\
\,
''
' ......\ "'"
TIME
FIG.9: PULSE WIDTH IS MEASURED at
the 50% marks on the waveform. As the
circuit's shunt capacitance is increased,
the rise time slows, causing the pulse to
become wider.
100
.01
0.1
1
10
~,...
100
1000
FREQUENCY I '1!Hz)
FIG.10: WHEN MEASURING SINE waves,
it's important to know how the probe's
input impedance (Zp) changes with regard
to frequency.
1000
'\
\.
100%
[\
G'
<
97%
'r-
"'
~
~ 100
'r-.
~
70.7%
3dB
e.
w
"'<
:;
0
>
10
1
10
30
100
FREQUENCY I MHz)
FIG.11: THE MAXIMUM VOLTAGE that a
probe can handle decreases as the
frequency increases.
inversely proportional to the frequency. Most scope probes are supplied with their own voltagederating specification.
Bandwidth
Scope probes are often rated for
bandwidth. It's best to use a probe
that has a bandwidth equal to or
higher than that of your scope. If
the probe's bandwidth is less than
that of the scope, then the input frequency will be limited by the probe.
Fig.12 shows the response curve
of a probe having a 100MHz bandwidth. By definition, bandwidth is
the upper frequency where the
scope's displayed voltage is down
8
SILICON CHIP
100
FREQUENCY 1MHz)
FIG.12: PROBE BANDWIDTH is the point
where the voltage amplitude is down 3dB
from a starting reference level.
3dB from the reference frequency
voltage. The formula to calculate
decibels is:
dB = 20 log Vout!Vin
For example, if the input is a 1V
p-p 100MHz sinewave, then at
- 3dB the scope 's waveform will
show an amplitude of only 0.707V
p-p (which is an amplitude accuracy of 70.7%). For an amplitude
accuracy of 90% or better, you
must limit the input frequency to
about 30MHz.
Another useful equation shows
how bandwidth is related to
risetime (tr):
Bandwidth = 0.35/tr
From this it follows that the
faster the risetime, the greater the
bandwidth.
For low-frequency applications
(audio frequencies), you can choose
a xl passive probe because it costs
the least and will do the job. But be
aware that the xl probe has a
limited bandwidth - less than
40MHz. On the other hand, use the
xlO probe for general digital,
analog and RF measurements. A
xlO probe has low capacitance and
a bandwidth upwards to 400MHz,
depending on the model and the
cable length.
Note also that if you are using
switched probes which are the most
commonly available, their stated
bandwidth only applies to the x10
mode. In the direct (xl) mode, their
bandwidth is much less. For example, the Coline SPlO0 probe has a
rated bandwidth from DC to
100MHz in the xlO mode but its
bandwidth drops to 10MHz in the
xl mode. Its capacitive loading also
increases from 16pF to 55pF (plus
the input capacitance of the
oscilloscope).
Although not discussed in this article, for frequencies higher than
400MHz choose one of the active
probes (FET) with a xl sensitivity.
This type of probe will provide high
sensitivity, low shunt capacitance
and a bandwidth greater than
900MHz.
Rp
CP
10M
11.2pf
GROUND LEAD L
FIG.13: GROUND-LEAD INDUCTANCE
will reduce the high-frequency
response through its series-inductive
reactance.
RINGING
,J
l\tWrG--.
Grounding
How often have you touched the
probe tip to an IC pin only to see a
waveform you know from experience isn't right? Then, after
moving the probe's grounding clip
from the chassis to the IC's ground
pin, the scope's trace immediately
shaped up and became recognisable as the waveform you've seen
before.
This leads us to the obvious question: how does the probe's grounding lead affect the circuit
measurement? The obvious answer
is that improper grounding will
generally distort the waveform by
allowing excess noise to be picked
up. That's true but it's only part of
the reason.
Fig.13 shows an equivalent circuit of a passive probe connected to
a voltage source. Notice the series
ground-lead reactance, L, whioh
represents the ground return path.
Rp and Cp represent the equivalent
impedance as seen at the compensated probe's tip. When measuring
any signal, the series inductive
reactance will be proportional to
both frequency and inductance by
the formula:
X1 = 2-irfL
The higher-frequency harmonics
will therefore see a larger inductive
reactance than the lower frequency harmonics. The pulse
waveform displayed on the
oscilloscope will show distortion
and aberrations because the p-p
voltages of the higher-frequency
FIG.14: PULSE RINGING OCCURS
when the ground-loop inductance and
probe tip capacitance form a series
resonant circuit that is shock-excited
by a very fast rise-time pulse.
harmonics have been attenuated
and phase shifted across the
ground-lead inductive reactance.
Now let's get back to the original
problem. If you move the probe's
grounding clip from the chassis to
the..IC itself, then the ground-loop
inductance will be reduced. That
allows the high-frequency harmonics to reach the scope's input,
so the trace shapes up. As a rule of
thumb, when making any kind of
measurement - such as amplitude,
rise time and pulse width - you
should use the shortest grounding
path possible.
As shown in Fig.14, loop inductance may also manifest itself as
ringing on the leading and trailing
edges of the signal pulse. The
ground-lead inductance and probetip capacitance form a seriesresonant circuit with only a lOMQ
resistor for damping. When shockexcited by a pulse, the resonant circuit will ring with a predictable
damped oscillation.
For example, an 11.2pF passive
probe having a 15cm ground lead
will ring at about 140MHz when hit
by a fast rise-time pulse. As the
ringing frequency increases, it
tends to fall outside the scope's
passband and is highly attenuated.
It's therefore desirable to try to increase the ringing frequency. To do
that, use the shortest possible
ground lead and the probe with the
lowest shunt capacitance.
It's ironic but for the reasons just
mentioned, you're more likely to see
ringing on an expensive high passband 300MHz scope than a low
passband 20MHz scope.
Mechanical properties
Often, while touching or rearranging a probe or lead wire, unpredictable, confusing and nonrepeatable effects are produced on
the observed waveform. That kind
of problem may have more to do
with the mechanical nature of the
probe than anything else.
If the probe's inner signalcarrying conductor is poorly shielded, then the probe's cabling will be
susceptible to external electric
fields. (The shielding could become
frayed due to constant flexing of
the cabling over many months or
years of use). A poorly shielded
wire can act as an antenna and
pick up all types of interference
such as electrical noise from
fluorescent lamps, radio stations
and signals generated by nearby
equipment. To virtually eliminate
any external field pickup, always
use a probe with coaxial cable
shielding of the centre conductor.
As a final note, the probe tip
should be clean or a poor circuit
connection will result. Also, be sure
to check printed circuit boards for
a conformal coating which is
sometimes used to guard against
humidity and static. This can easily
prevent an electrical connection
between the probe tip and the circuit. It may be necessary to scrape
off some of that coating to make a
good connection to the circuit. ~
Footnote: next month, we plan to
describe a practical x10 CAO
probe that you can build yourself.
This unit is essentially a practical
version of the circuit shown in
Fig.5 on page 6.
JUNE 1989
9
|