This is only a preview of the August 1996 issue of Silicon Chip. You can view 24 of the 96 pages in the full issue, including the advertisments. For full access, purchase the issue for $10.00 or subscribe for access to the latest issues. Items relevant to "Electronic Starter For Fluorescent Lights":
Items relevant to "Build A VGA Digital Oscilloscope; Pt.2":
Articles in this series:
Articles in this series:
Articles in this series:
Articles in this series:
|
The oscilloscope is a wonderful measurement
tool but if it is not used carefully it can give
highly misleading results. You can achieve the
full potential of your scope but only if you know
what you are doing. This article gives some good
tips on oscilloscope use.
By BRYAN MAHER
Say you have invested hard cash in
a good quality oscilloscope. It looks
a beautiful instrument and the specs
guarantee it to be accurate within
2%. Wow! And its bandwidth is wide
enough to make your friends drool.
But a scope is only a tool, no matter
how glossy the literature. If you don’t
use it properly you will be disappointed with the results.
Let’s start with a simple DC measurement, using the circuit shown in
Fig.1(a). If we read the DC voltage at
point D, a digital voltmeter (DVM)
gives a reading of +4.9V. If we then
connect the oscilloscope via a 1x
shielded probe, the deflection on the
screen is likely to indicate only about
+4.17V. Which is correct? Clearly that
64 Silicon Chip
scope probe is loading the source of
this measurement, pulling the voltage
down!
“Source” here means any part of
circuit at which we make a measurement. In this case it is point D in Fig.1.
And “source resistance” or “output
resistance”, denoted by Rs, means the
ratio of the change in voltage at that
point (caused by attaching the probe)
divided by the minute current drawn
by the probe.
This is denoted by the expression:
Rs = (∆v/∆i) Ω.
Because it is a voltage/current ratio, we call it resistance (ohms), even
though it is a calculated quantity.
Only rarely is Rs a single physical
component. Nevertheless Rs does have
the ability to upset the workings of a
circuit. “Delta” simply means a small
change in any quantity.
Equivalent circuit
The equivalent circuit, illustrated
in Fig.1(b), reveals how this loading
effect occurs. The input resistance of
the direct 1x probe connection is just
the 1MΩ resistor within the scope,
which we have called R1 in Fig.1(a).
R1 and Rs actually form a voltage
divider, so the scope sees only the
voltage at D, which is the true voltage
of the source reduced by the fraction
(R1/(R1 + Rs)).
Typically, a digital multimeter has
an input resistance of 10MΩ so using
it has a less deleterious effect on the
voltage. This is why the DMM reading
is higher, at +4.9V.
You can calculate the value of the
source resistance Rs in this case from
these measurements and the definition
given above. It works out to be about
200kΩ which is reasonable for this
particular op amp circuit.
Let’s define V as the unloaded
output voltage of the source; ie, the
potential at point D when neither the
scope probe nor the DMM is connected
to it. Using the voltage divider equa
tion, the voltage Vpat D when only the
1x probe and scope is hooked on is:
Vp = V(R1/R1 + Rs)
= V(1MΩ/1.2MΩ)
= V/1.2
The scope reads Vp as +4.17V, so
the unloaded output vol
tage at the
point D is:
V = (1.2)(4.17) = 5V
The relatively low resistance of the
scope input was the cause of the loading effect. It loaded the source and so
caused the oscilloscope to read +4.17V
instead of the true +5V.
Measurement rule-of-thumb
The cure for this loading effect is
now obvious. The test instrument
should have an input resistance much
greater (preferably 100 times greater)
than the output impedance of the
source to be measured. A 100 times
factor would limit loading errors to
about 1%. But practical aspects like
price, availability and frequency
response will limit our selection of
scope probes.
A common favourite, the 10x probe,
as illustrated in Fig.2, is an excellent
choice in most cases. This type of
probe contains a 9MΩ resistor called
Rp. Therefore the total probe connection resistance, Rin, is equal to Rp in
series with the scope input resistance,
R1. That is:
Rin
= (Rp + R1)
= (9MΩ + 1MΩ) = 10MΩ
If we substitute this 10x probe in
the measurement shown in Fig.1, the
oscilloscope would display a deflection of +4.9V, the same as the DMM
reading, a satisfying result.
Fig.1: this dual phase amplifier (a) has a 5V output at point D where the
source resistance is 200kΩ. But clipping the 1MΩ probe onto this point pulls
the voltage down to 4.17V. The equivalent circuit (b) shows that the source
resistance Rs forms an unwanted voltage divider with R1, the input resistance
of the 1x probe and the scope. This reduces the voltage seen by the scope.
High voltage measurements
Fig.2 shows a second important
use of the 10x probe. Here the source
resistance is quite low (due to negative
feedback) at the collector of transistor
Q1 so loading is not a worry but the
high voltages are! In this case we can
use the fact that Rp (in the probe head)
and R1 (in the oscilloscope) form a
deliberate voltage divider. Any voltage
which we apply to the probe tip will
be reduced at the scope input terminal.
The reduction fraction is:
Vsc = R1/(Rp + R1)
= 1MΩ/(9MΩ + 1MΩ)
= 1/10.
That’s why this probe is known as a
10x, because it produces a 10:1 voltage
attenuation.
In the circuit of Fig.2, the high volt-
Fig.2: the 9MΩ probe resistor Rp and the 1MΩ scope input resistor R1, form a
deliberate voltage divider. This reduces the voltage at the oscilloscope input
terminal to one tenth of that at the probe tip.
age of the supply (+450V) rules out
use of the 1x probe and forbids direct
connection to the scope’s input. But
the 10x probe is suitable, provided it
has a voltage rating above 450V. This
probe will reduce all waveform voltages to one tenth and the DC voltage
at the scope input will be no more
than +45V.
By dividing down the signal, the
10x probe effectively multiplies the
V/div calibration on the attenuator
switch by a factor of 10. So a 5V/div
setting now means 50V/div and eight
vertical divisions on the screen will
correspond to a 400V range. Hence
this 360V signal fits within the graticule limits.
Many top line scopes can sense
when the 10x probe is connected to the
modified BNC input terminals. Then
internal logic circuits multiply the
August 1996 65
Fig.3: source (a) has output resistance Rs equal to 50Ω at point D.
The high frequency equivalent circuit (b) shows that Cp forms an
unwanted voltage divider with Rs. Cx represents the combined
stray capacitance of the coaxial cable and the scope input.
10x probe) we must be aware that the
probe tip still carries a lethal 360V! For
safety we must keep the amplifier 0V
line connected to the scope frame and
to mains earth. And we never unplug
the probe from the scope while the
probe tip is still hooked onto a high
voltage point.
All probes which contain only resistors and capacitors are called passive and oscilloscope manufacturers
market a range of higher resistance
units. A few of these are listed in Table 1 but not all probes on the market
have voltage ratings as high as those
shown here.
Direct 1x scope probes have only a
small series resistance so they cause
little attenuation of the signal being
measured. They are useful for the
display of very small voltages of low
frequency signals, when measured at
low impedance points, such as the
outputs of op amps.
Some less common sources, like
biological assay electrodes, have an
extremely high output resistance.
To display signals from these, active
probes are required. Typically, these
employ IGFETs and other active circuitry to provide an input impedance
of 10GΩ and zero input capacitance.
Oscilloscope bandwidth
Fig.4: the amplitude response of an oscilloscope falls at high frequencies. At
full rated bandwidth, the response is -3dB or 30% lower than it is at low
frequencies.
on-screen readout by 10, to correctly
display the voltage value at the probe
tip. This facility is not provided in
cheaper scopes and nor does it work
when a scope is used with a probe of
a different brand.
Safety precaution
Though the oscilloscope is safely
working on reduced input voltages
(because of the attenuation by the
Fig.5: With AC (capacitive) coupling,
the signal passes through a high pass
filter. This will reduce the amplitude
of low frequency signals and distort
low frequency pulse waveforms.
Table 1
Probe
Attenuation
1x
10x
100x
1000x
R(in)
1M
10M
10M
100M
66 Silicon Chip
Maximum DC
Voltage
350V
600V
1.5kV
20kV
Derated Above
Derated to
1MHz
200kHz
100kHz
30V <at> 20MHz
300V <at> 20MHz
2kV <at> 20MHz
Another scope parameter which
new users often have difficulty coming to terms with is bandwidth. This
could be easily measured if you had a
synthesised RF signal generator with
an output of 5V over a frequency range
from 100kHz to 250MHz and an output
impedance of just 50Ω.
You might think that such a wide
band source could easily demonstrate
a scope’s bandwidth. Would you just
connect the 10x probe to the generator
and then sweep over the frequency
range? Fig.3 illustrates the setup, with
the probe’s internal resistance and
capacitance shown.
However, you might be disappointed to find that, when the generator
was set to the advertised bandwidth
frequency of your high performance
scope, say 250MHz, the vertical deflection is only half what it should
be. So what does scope bandwidth
mean?
The bandwidth of any oscilloscope
is that high frequency at which the
response has fallen to 70.7% (-3dB),
compared to the reference frequency
value, as illustrated in Fig.4. This
Table 2
Taken from a Tektronix TDS360 digital oscilloscope, this screen printout
shows the effects of incorrect adjustment of 10x probes on the scope’s internal
1kHz compensation signal. Channel 1, the upper trace, shows too much probe
capacitance (over-compensation) while the channel 2, lower trace, shows
insufficient capacitance (under-compensation). The correct probe compensation
adjustment would show a square wave with “square” corners.
This scope printout shows the effects of DC and AC coupling on a pulse
waveform with uneven duty cycle. Channel 1, top trace, is DC coupled and it
can be seen that the voltage swings equally above and below the zero reference
line (solid horizontal cursor). Channel 2, lower trace, is AC coupled and the
waveform has floated down with respect to the zero reference line (dotted
horizontal cursor).
shows that the response of any oscilloscope is down by 30% at its advertised
full bandwidth!
Furthermore, Fig.4 shows that the
manufacturer’s guarantee of an amplitude error of less than 2% only applies
for signal frequencies less than one
quarter of the rated bandwidth.
Frequency
Capacitive
Resistance
1MHz
10MHz
50MHz
100MHz
250MHz
300MHz
400MHz
13.3k
1.3k
265
132
53
44
33
Therefore, to make amplitude
measurements with less than 2%
error, we need a scope with a quoted
bandwidth four or five times higher
than the signal frequency. For example, accurate amplitude display of a
50MHz sinewave requires a 250MHz
oscilloscope.
This is only part of the bandwidth
story. As we noted above, testing an oscilloscope with a wideband generator
could show an error of more than 50%
at the advertised scope bandwidth.
How could it get worse?
In most cases the advertised -3dB
bandwidth of a scope applies only
when signals are coupled directly into
the instrument front terminal and not
via a probe, because probes also have
frequency limitations.
This is demonstrated by Fig.3(b),
which is the high frequency equivalent
of the circuit shown in Fig.3(a). As
before, the resistance presented by the
probe and scope connection is:
Rin = (Rp + R1) = 10MΩ
where Rp is the resistance inside the
probe and R1 is the input resistance
of the oscilloscope.
In the equivalent circuit of Fig.3(b)
we can ignore the 10MΩ input resistance Rin because it is so much higher
than the 50Ω source resistance Rs. But
we cannot discount the probe’s input
capacitance Cp which is equal to 12pF.
The capacitive reactance of Cp is:
Xc = 1/(2πfCp).
This forms an unwanted voltage
divider with the source resistance Rs.
At high frequencies the resulting low
value of Xc drastically reduces the
signal amplitude before it enters the
scope. Table 2 demonstrates the severity of this effect, with the reactance of
12pF at specific frequencies.
From Table 2, we observe that at
250MHz the probe’s capacitive reactance has fallen to 53Ω. Now we will
August 1996 67
Fig.6: since a PWM signal has a varying duty cycle and therefore an effectively
varying positive and negative DC offset, AC coupling will cause the waveform to
waver above and below the 0V reference line.
see the reason why the amplitude
displayed on the screen fell to 50%.
Firstly, looking at Fig.4(b), we see
that at 250MHz the voltage divider
effect of the 53Ω Xc with the 50Ω
source resistance Rs reduces the
signal voltage at D down to 70% of
the unloaded source voltage (it’s a
vector calculation, because of the
capacitor).
Secondly, as Fig.4 shows, the
displayed amplitude will be further
reduced to 70% of the voltage at the
scope input, because the signal frequency is now equal to the 250MHz
bandwidth of the scope. So the amplitude you would see on the screen
will be reduced to (70% x 70%) =
50% of the unloaded source voltage.
That explains why a high frequency
measurement with a 10x probe can
have such large errors.
Table 3
Attenuation
R(in)
C(in)
1x
10x
10x
100x
10x
10x
1M passive
10M passive
10M passive
10M passive
100k active
500 divider
55pF
12pF
8pF
2.7pF
0.4pF
0.15pF
68 Silicon Chip
Only in a few cases will a manufacturer guarantee that the advertised
bandwidth applies at a specified probe
tip. Examples include the Tektronix
400MHz oscilloscope model 2465B
but only when used with their 1MΩ
passive 10x probe model P6137.
Table 3 shows the input capacitance
and bandwidth of typical probes.
Frequency pulling
Often, the application of a passive
scope probe to some points of a circuit
can have drastic effects, particularly
in the case of crystal and other oscillators.
These require critical positive feedback gain and phase, set by specific
small capacitor values, to maintain oscillation at the required frequency. But
hooking a passive probe onto a high
impedance point of these circuits can
add 12pF of capacitance, upsetting the
feedback. This action
can either reduce the
Bandwidth
operating frequency
or may stop oscilla15MHz
tion altogether.
100MHz
How do we avoid
500MHz
this? Many systems,
250MHz
including some TV
4GHz
rec eivers, contain
buffered test points,
9GHz
where sensitive circuits are accessed
either via an inbuilt resistor or a low
impedance source follower.
Alternatively, a simple expedient is
to attach a small resistor, about 10kΩ,
to the probe and use the other end of
that resistor as the probe point. The results may be inaccurate but at least you
can monitor the waveforms. Another
alternative is to use a high impedance
active probe, such as listed in Table 3.
For frequencies above 500MHz,
wideband active FET probes are
available with a high input impedance
and they require a separate supply.
Examples include the Tektronix type
P6204 which has a 1GHz bandwidth
and the type P6217 which operates
to 4GHz. Active probes accept small
input voltages, typically below 10V.
For really wide bandwidth scopes,
between 2GHz and 10GHz, low impedance divider probes are available, with
input resistances of 50Ω, 500Ω or 5kΩ.
They plug into the 50Ω input terminals
on very high frequency oscilloscopes.
Probe risetimes
Another area where a new oscilloscope can disappoint is when displaying square waves which are supposed
to have fast rise and fall times. Fig.3(a)
shows the connection as before and
now we will explain why the probe
capacitor Cp is there at all, in view of
the trouble it causes when displaying
very high frequencies.
The reason why Cp is inside the
and undershoot. Naturally this Cp
adjustment also has a big effect on the
displayed bandwidth so if you don’t
adjust it correctly, it is yet another
source of measurement error.
AC coupling
This scope printout shows the effects of DC and AC coupling on a pulse
waveform with varying pulse width (ie, pulse width modulation). Channel 1,
top trace, is DC coupled while Channel 2, the lower trace, is AC coupled. The
varying pulse width effectively becomes a varying DC offset which is reflected
as a wavy modulation on the waveform, an erroneous display. This is the same
effect as depicted in Fig.6.
probe becomes clear when you look at
pulse risetimes. The probe’s shielded
cable and the oscilloscope’s input
stage add up to a considerable capacitance to ground, probably between
35pF and 100pF.
This we denote as Cx in Fig.3(a).
If Cp did not exist in the probe head,
then the probe resistor Rp, together
with this stray capacitance Cx, would
form a severe low pass filter. The
effect would be a reduction in amplitude and a phase change in sinewave
signals and a drastic slowing of the
risetime of pulses as displayed on
the screen.
Therefore the capacitor Cp has been
deliberately included in the probe to
correct these errors. But Cp must be
correctly adjusted until the two time
constants, RpCp and R1Cx, are equal.
To facilitate this adjustment, most
oscilloscopes provide a fast-risetime
1kHz square wave calibrating signal
from a terminal (usually) on the front
panel. You just hook the probe onto
this CAL terminal and adjust the probe
capacitor Cp until the scope displays
a true square wave.
If Cp is set too low, the square wave
will be rounded off while if Cp is too
high, the square wave will overshoot
So far we have talked about large
DC voltages and high frequencies but
if you have a circuit with high DC
voltages and small signals, you need
to switch the scope’s input to AC cou
pling. This enables you to use high
input sensitivity while blocking out
a large quiescent DC voltage. As Fig.5
illustrates, the signals then must pass
through the R1C1 time-constant. This
will reduce the amplitude of low frequency signals, distort square waves
and pulses and can play merry hell
with pulse width modulation (PWM)
signals.
To see why, we need to critically
look at just what it means to feed a
signal through a coupling capacitor.
In Fig.6 we have sketched a PWM
signal which is applied to the left
side of capacitor C1. Below that is the
waveform which appears on the right
hand side of C1 and is displayed on
the oscilloscope screen.
At time t7, the input signal lifts
the left side of C1 from zero to +10V,
charging the capacitor. So the right
side also rises to +10V. Between times
t7 and t8, the input voltage remains
steady. But the charge on C1 leaks
away through the resistor R1, lowering
the voltage on the right hand side of
the capacitor from +10V to +8V. Then
at time t8, the input voltage drops from
+10V to zero.
Because this fall is abrupt, the potential on the right side of the capacitor
must also fall by 10V; ie, from +8V to
Fig.7: one possible circuit
for the Chop/Alternate
section of an analog
scope. CMOS analog
switches alternately
switch the signals from
channels 1 and 2 through
to the vertical deflection
amplifier.
August 1996 69
Fig.8: this series of waveforms illustrates how the Chop mode in an oscilloscope
rapidly chops between the input channels to produce two waveforms on the
screen. Waveforms (c), (d) and (e) show an expansion of the 1ms period in
waveforms (a) and (b).
-2V, taking the displayed trace into
negative regions. This may leak away
to about -1.7V by time t9, when the
input rises again. This time the +10V
change in the input signal lifts the
display up to +8.3V. During the long
constant input between t9 and t10, the
display again leaks down to +6.6V. You
can see that the displayed waveform
is far from the truth.
When the positive input pulses
are long, with a duty cycle greater
than 50%, the display progressively
migrates downwards (duty cycle is
the ratio of the pulse on-time to the
pulse off-time). If the duty cycle remained constant, after many cycles the
displayed signal would be displaced
until the area enclosed between the
positive regions of the waveform and
the zero line is equal to the area enclosed between the negative regions
and the zero line.
70 Silicon Chip
By this rule, the long duty cycle
between times t7 and t12 will push the
waveform downwards. But the same
rule means that between times t13 and
t16, when the duty cycle is short, the
waveform display must rise above the
zero line, in order to equalise positive
and negative areas.
So the complex PWM waveform of
Fig.6 will rise and fall as the duty cycle
changes. The only cure is to monitor
the waveform with DC coupling. AC
coupling is a trap for young players –
use it only when you must block high
DC voltages.
Dual-trace operation
One of the really powerful benefits
of a scope is the ability to monitor two
signals at once but here again there
are traps. If you want to measure the
timing or phase differences between
two signals you need to know just
how your scope displays two different
inputs on the screen simultaneously. What we are talking about is the
choice between Alternate and Chop
modes.
Fig.7(a) illustrates one possible
circuit for the Chop/Alternate section
of an analog scope. Two different signals on channels 1 and 2 firstly pass
through their individual attenuators
and preamplifier stages A1 and A2,
then to the Chop/Alter
nate section
which includes IC1, IC2 and IC3. You
will easily follow its operation as we
view it a bit at a time.
IC1a, b and d are CMOS analog
switches and each turns on only when
a logic high signal is applied to its
control terminal. For example, IC1a
conducts between pins 4 and 3 only
when a logic high is applied to pin 5.
The timebase section of the oscilloscope, as well as providing the
horizontal sweep, also feeds a control
signal in at point T. This controls all
four CMOS switches via inverters
IC2a, IC2b and IC2c. IC3 is a summing
operational amplifier, while Ri1 and
Ri2 are its two input resistors and Rf
is the feedback resistor. Point X is the
summing junction.
The gain from either channel 1
or channel 2 inputs to the output at
point N is -(Rf/Ri) = -(10kΩ/10kΩ) =
-1. Signals from point N feed to the
vertical deflection amplifier for display on screen.
Now what happens when we select
the Alternate display mode? Say we
apply a signal to channel 1 input and
a square wave to channel 2. If the timebase section feeds a low control signal
to the point T, this will be inverted in
IC2b and will present a logic high to
pin 5 of gate IC1a, turning it on.
So the sinewave signal on channel
1 will feed through A1, through Ri1,
IC1a and IC3, and will pass on to the
vertical deflection amplifier, to be displayed on the screen. At the same time,
gate IC1b is off, so channel 2 signals
cannot pass to the vertical deflection
amplifier.
But when a logic high signal is fed to
point T, the conditions reverse. Analog
switches IC1b and IC1c will conduct
and IC1a turn off, allowing the channel
2 signal to be displayed on the screen.
The control signal at point T is high on
the 1st, 3rd, 5th, 7th, etc sweeps and
low on the 2nd, 4th, 6th, 8th, etc. Thus,
all odd sweeps display the sinewave
on channel 1 and all even sweeps show
the square wave on channel 2.
You can use the individual vertical
position (shift) controls to move the
two displays apart. At slow sweep
speeds, the display alternates between
signal 1 in the upper half of the screen
and signal 2 in the lower screen.
At fast sweep speeds, the persistence of the screen phosphor enables
you to see both signals continually on
the screen. Hence, Alternate mode is
successful with fast sweep speeds but
unsuitable at slow sweeps.
Chop mode
Now what happens if you change to
“Chop” mode. This causes a separate
high frequency oscillator within the
timebase unit to toggle the control
signal fed to point T (toggle means to
switch continually between logic high
and low). This is done at a fixed fast
rate, perhaps 10kHz, as illustrated in
Fig.8 but in some high frequency osc
illoscopes the toggle rate may be as
high as 1MHz.
In the example shown in Fig.8, the
main sweep is switched to 100 mil-
liseconds per division, which takes
one second for each full sweep. The
sinewave on channel 1 has a frequency
of about 3.5Hz and the square wave on
channel 2 is about 6Hz. The control
signal at point T has period equal to
1/10kHz = 100µs as Fig.8 shows.
This makes channel 1 conduct
through IC1a for 50µs, channel 2 conducts through IC1b for the next 50µs
and so on. Both input signals are thus
chopped up into thousands of little
time segments 50µs long, like two lines
of ants crawling across a page.
On the screen are displayed these
20,000 discontinuous segments of
the input signals, as IC1a and IC1b
conduct in turn. A small sector of both
traces is shown in Fig.8(d) & (e) drawn
one thousand times time-expanded.
While T is at logic high, a small segment of the sinewave (a) is displayed
in the upper half of the screen. But
when T is at logic low, a short piece
of the square wave (b) appears on the
lower half at (e). While one signal is
displayed, the other is blanked off.
This process continues repeatedly,
right across the screen. The slight
blur
ring due to the width of the
light spot makes each trace appear
continuous.
If we raised the sweep speed sufficiently we would see the discontinuous nature of the display. So chop
mode is unsuitable for very fast sweep
speeds. In some scopes Chop mode is
automatically selected at slow time
base speeds and Alternate is selected
at high sweep speeds.
Now we can see why Chop and
Alternate modes can affect timing
and phase comparisons between two
different signals. Alternate mode leads
to impossibly wrong results, because
it allows the oscilloscope to trigger
independently on each channel; time
correlation is completely lost. Therefore, Chop mode must be used when
comparing the timing of different
signals.
Phase shift
A final vital point to note here is
the phase shift which AC coupling
produces, as noted above. Therefore,
when comparing phases and timing
of different signals, switch both channels to DC coupling or switch both
channels to AC coupling. Don’t have
channel 1 AC-coupled and channel 2
DC-coupled; that will lead to serious
SC
errors.
YOU CAN
AFFORD
AN INTERNATIONAL
SATELLITE TV
SYSTEM
SATELLITE ENTHUSIASTS
STARTER KIT
YOUR OWN INTERNATIONAL
SYSTEM FROM ONLY:
FREE RECEPTION FROM
Asiasat II, Gorizont, Palapa,
Panamsat, Intelsat
HERE'S WHAT YOU GET:
●
●
●
●
●
●
400 channel dual input receiver
preprogrammed for all viewable satellites
1.8m solid ground mount dish
20°K LNBF
25m coaxial cable
easy set up instructions
regular customer newsletters
BEWARE OF IMITATORS
Direct Importer: AV-COMM PTY. LTD.
PO BOX 225, Balgowlah NSW 2093
Tel: (02) 9949 7417 / 9948 2667
Fax: (02) 9949 7095
VISIT OUR INTERNET SITE http://www.avcomm.com.au
YES GARRY, please send me more
information on international band
satellite systems.
Name: __________________________________
Address: ________________________________
____________________P'code:
__________
Phone: (_______) ________________________
ACN 002 174 478
August 1996 71
|