Three-Way
USB Scope
Shootout
. . . are they any good?
Are you curious about those low-cost PC-based digital sampling
oscilloscopes currently available via the web? This comparison
review of the three most common units shows that they can be a
cheap alternative to a full-size scope.
By JIM ROWE
U
SING YOUR LAPTOP PC as the
heart of a digital oscilloscope has
a lot of appeal, as most laptops have a
much larger screen than that in most
free-standing digital oscilloscopes. But
are USB scopes any good?
Typically, the vendors all claim
outstanding performance, yet they are
all overseas and there’s no chance of
being able to try their units out; you
70 Silicon Chip
have to buy them sight unseen.
What about reviews in electronics
publications? I could find only one
solitary review, of the Hantek DSO2250 USB, written by Geoff Graham,
a frequent contributor to SILICON CHIP,
on his website at http://geoffg.net
The Hantek scope has been available
for about six years – under a variety of
other brand names including Protek,
Acetech, Voltcraft and Tenma. Geoff
Graham didn’t rate the DSO-2250 very
highly and a more recent online video
review really tore it to shreds. It is
included in this comparison because
it is well-known. The other two units
are the Virtins DSO-2820R and the
Link Instruments MSO-19.2 (which is
actually a mixed signal scope).
Hantek and Virtins’ units are made
siliconchip.com.au
The Hantek DSO-2250 comes with two switchable (10x/1x) 100MHz probes, a USB cable, an 80mm software CD and a
15-page user manual. It has a maximum sampling rate of 250MSa/s and an analog bandwidth of 100MHz, although the
sampling rate drops to 125MSa/s when both channels are being used (indicating that the sampling is interleaved).
in China, while the Link MSO-19.2
seems to be made in the USA (Fairfield, New Jersey). All three come with
drivers and software for PCs running
Windows.
Hantek DSO-2250
The Hantek unit is housed in a
moulded plastic box measuring 190
x 100 x 33mm and weighing 304g. It
comes with two switchable (10x/1x)
100MHz probes, a USB cable to be
hooked up to two USB ports on your
PC, an 80mm software CD and a
15-page A5-size user manual which
mainly covers software installation.
Hantek claims that the DSO-2250
has a maximum sampling rate of
250MSa/s and an analog bandwidth
of 100MHz. However, that only applies when a single channel is being
used. With both channels in use, the
sampling rate drops to 125MSa/s – so
the sampling must be interleaved.
Since the effective analog bandwidth is usually defined as the maximum sampling rate divided by at least
2.5, this means that when both channels are used the effective bandwidth
for each channel must drop to 50MHz.
This is still quite respectable, although
it’s not necessarily achieved by the
DSO-2250’s analog input channels.
siliconchip.com.au
When it comes to the size of the internal buffer memory, Hantek specifies
a range of 10K – 512K sample points.
This upper limit sounds good but after
playing with their software for quite
a while I still couldn’t figure out (a)
whether these figures apply to one or
both channels or (b) how to increase
the buffer size from 10K anyway. This
figure seems to be fixed, although it
does seem to apply to each channel.
For triggering, the DSO-2250 software lets you choose either one of
the two main input channels, both
alternately, or the external trigger input
(either directly or via a 10:1 attenuator).
It also provides a choice of Auto/Normal/Single shot trigger modes, edge
triggering (± slope), automatic 50%
level triggering and adjustable triggering level over a range of ±4 full vertical
divisions. Plus you can also select an
HF rejection filter.
There’s quite an impressive range of
waveform voltage and time/frequency
measurements: Vpp, Vmax, Vmin,
Vmean, Vrms, Vamp and ±overshoot;
plus period, frequency, ±pulse width
and rise/fall times. You can also average between 1-128 traces for noise
reduction, select either Step, Linear
or Sin(x)/x interpolation and even
select a virtual display persistence
of either zero or within the range of
100ms – 10s.
In addition, the DSO-2250 provides
an FFT (Fast Fourier transform) based
Spectrum Analyser function, with five
selectable bin sizes (256, 512, 1024,
2048 or 4096) and a choice of Rectangular, Hanning, Hamming or Blackman
window functions.
Hantek doesn’t specify the current
drawn by the DSO-2250 from the PC’s
USB port but the fact that it is supplied with a USB cable fitted with two
Type-A plugs at the PC end suggests
that its current drain is fairly hefty.
We measured it at between 560mA
and 580mA. That explains the second
Type-A plug, to share the load between
two of the PC’s USB ports.
Virtins DSO-2820R
We reviewed the Virtins Multi-Instrument virtual instrument software
MI3.2 in the September 2012 issue
of SILICON CHIP. This software was
capable of turning a PC with a decentquality sound card into a 2-channel
audio scope combined with an FFT
spectrum analyser, plus a 2-channel
audio signal/function generator.
In the same issue, we described a
Virtual Instrument Audio Test Interface
and you can see a 2-page preview at
February 2015 71
Virtins’ DSO-2820R is a 2-channel USB DSO and FFT spectrum analyser, with a maximum sample rate of 200MSa/s (one
channel) or 100MSa/s (two channels) and a claimed analog bandwidth of 80MHz. It’s housed in an extruded aluminium
case and comes with a pair of switchable (10x/1x) 100MHz test probes, a USB cable and a CD with Virtins MI software, the
Windows drivers and two manuals in PDF format.
The Link Instruments MSO-19.2 is the smallest of the three units but still packs in a single-channel DSO with a maximum
sampling rate of 200MSa/s and a claimed analog bandwidth of 60MHz. Also included is an 8-bit digital logic analyser
that can also sample at up to 200MSa/s (synchronised with the DSO) with decoders for SPI and I2C serial digital signals, a
100MSa/s pattern generator or digital word generator, and an FFT spectrum analyser covering from DC to 100MHz. And it
also has the makings of a pulse-type time domain reflectometer or TDR!
www.siliconchip.com.au/Issue/2012/
September/USB+Test+Instrument+In
terface+For+PCs
The DSO-2820R is one of a number
of PC-based DSOs from Virtins now,
all supplied with the Multi-Instrument
software – now revised as MI3.4.
The DSO-2820R is again a 2-channel USB DSO and FFT spectrum analyser, with a maximum sample rate of
200MSa/s (one channel) or 100MSa/s
(two channels) and a claimed analog
bandwidth of 80MHz. (There’s also the
72 Silicon Chip
lower cost DSO-2810R [US$200] with
a maximum sample rate of 100MSa/s,
and the larger and more expensive
DSO-2820E [US$370] which combines
the 2820R DSO and spectrum analyser
with a 10-bit 200MSa/s DC-60MHz
arbitrary waveform/signal generator.)
Smaller than the Hantek unit, the
Virtins DSO-2820R is housed in an
extruded aluminium case measuring
115 x 70 x 25mm and has an overall
weight of 150g. It comes with a pair
of switchable (10x/1x) 100MHz test
probes, a USB cable to hook it up to
one USB port on your PC and a 120mm
CD with a bundled copy of Virtins MI
software, plus the Windows drivers
and two manuals in PDF form – ie,
an 80-page hardware manual and a
296-page MI software manual (some
of which is not applicable with the
DSO-2820R).
Virtins claims a maximum sampling
rate of 200MSa/s and an analog bandwidth of 80MHz for the DSO-2820R.
You will find that the 200MSa/s samsiliconchip.com.au
siliconchip.com.au
$249.00
$249.00
$220.00
COST IN
US DOLLARS,
(EXCLUDING
FREIGHT)
1x Analog Input
(1M//15pF),
8 x Digital
Inputs/Outputs
(Logic Analyser,
Pattern Gen),
1 x Trigger
Output
60MHz
8~16 bits/ch,
200MSa/s
(1Channel)
100MSa/s
(2 Channels)
Frame Mode,
2GSa/s
RIS Mode*
Sampling
8 bits (DSO),
200MSa/s
Single Shot,
2GSa/s
RIS Mode*
Sampling
2 x 80MHz
8 bits/channel,
250MSa/s
(1 Channel)
125MSa/s
(2 Channels)
2x Analog
Inputs
(1M//25pF)
one External.
Trigger Input,
1 x Digital
(Probe Comp)
Output
2x Analog
Inputs
(1M//15pF),
1 x Digital
(Probe Comp)
Output
100MHz
(1 Ch),
2 x 60MHz
(2 Ch)
RESOLUTION,
MAXIMUM
SAMPLING
RATE
NUMBER OF
INPUT, OUTPUT
CHANNELS
CLAIMED
ANALOG
BANDWIDTH
1000 points
per channel
(DSO, Logic
Analyser)
In Streaming
Mode, Limited
only by PC
memory or
Hard Disk
In Frame Mode,
20K (16b) or 40K
(8b) Samples
per channel.
10K points
per channel
FRAME
BUFFER SIZE
92x61x20mm,
plastic (ABS),
74g
115x70x25mm,
metal
(Aluminium),
150g
190x100x33mm,
Plastic (ABS),
304g
PHYSICAL
CASE SIZE,
MATERIAL
& WEIGHT
1 x 1M/10M
(x1/x10)
Input Probe,
harness with
15x Digital
I/O leads,
9 x nano clips,
USB cable
(Type A plug
to Mini
Type B plug)
2 x 1M/10M
(x1/x10)
Input probes,
1 x USB cable
(Type A plug to
Type B plug)
2 x 1M (x1)/
10M (x10)
Input probes,
1 x USB cable
(2 x Type A
plugs, 1 x
Type B plug)
PROBES &
CABLES
SUPPLIED
YES
(120mm CD)
YES: Virtins
MultiInstrument
3.3
(120mm CD)
YES
(80mm CD)
WINDOWS
DRIVER &
SOFTWARE
SUPPLIED?
4-page A5
Install Guide
plus 100-page
Software
Manual in
PDF form
on CD
80-page A4
Hardware
Manual
Plus 296-page
M-I 3.3
Software
Manual
(both in PDF
form on CD)
15-page A5
User
Manual
only
HARDWARE &
SOFTWARE
MANUALS
SUPPLIED?
CURRENT
DRAIN
FROM PC’s
USB PORT(S)
560-580mA
running or
stopped
(2 USB ports
required)
322mA
running,
250mA
stopped
300mA
running,
275mA
stopped
TRIGGERING
FACILITIES
SOURCES:
Ch1,Ch2, Alt,
External, Ext/10
MODES:
Auto, Normal,
Single
Edge, ±Slope,
HF reject
TRIG LEVEL:
50% or full
scale adjustable
SOURCES:
ChA, ChB, Alt
MODES:
Auto, Normal,
Single, Slow
Edge, ± Slope,
HF reject, noise rej
TRIG LEVEL:
Full scale adj,
Adj Pre/Post
Dig Triggering
SOURCES:
DSO or any
combn of Logic
Analyser inputs
MODES:
Auto, Normal,
Single
Edge, ±Slope,
<pulse width>
TRIG LEVEL:
full scale adjust,
Pre/Post trig for
Logic Analyser
MAIN
FEATURES
ADDITIONAL
FUNCTIONS
Logic Analyser,
SPI and I2C
analysis
MEASUREMENTS:
Vp-p, Vmax, Vmin,
Vmean, Vmedian,
Vrms, VcursorA,
VcursorB
Period, Frequency,
±pulse width,
rise/fall time etc.
MEASUREMENTS:
Vmax, Vmin,
Vmean, Vrms,
dBV, dBu, dB,
dB(A/B/C)
Frequency, RPM,
duty cycle etc
(Many of these
via the ‘DMM’
display.)
SPI & I2C DECODERS
PATTERN GENERATOR:
Maximum 1023 points,
Rate 10kSa – 100MSa/s
8-bit LOGIC ANALYSER:
100MHz bandwidth
FFT/SPECTRUM ANALYSIS:
1000 points/channel
Seven Window Functions
CALIBRATION SIGNAL
GENERATOR:
Square wave or MLS,
Frequency 25MHz/N
where N = 1 – 25000.
FFT/SPECTRUM ANALYSIS:
Nine bin sizes
(128 – 32768),
50 Window Functions
MEASUREMENTS:
Vp-p, Vmax, Vmin,
Vmean, Vrms, Vamp,
± overshoot,
FFT/SPECTRUM ANALYSIS:
Period, Frequency,
Five bin sizes
± pulse width,
(256 – 4096),
rise/fall time etc.
Four Window Functions
Averaging
(Rectangular, Hanning,
(1�––128 traces
Hamming, Blackman)
Interpolation (Step,
Linear, sin(x)/x)
Persistence (Min,
100ms – 10s)
FIG.1: A COMPARISON CHART OF THREE LOW COST PC-BASED/USB POWERED DSO & MSO DEVICES
*RIS mode is ‘Random Interleave Sampling’, one variant of Equivalent Time Sampling or ‘ETS’ (used to achieve higher effective sampling rates, but for repetitive signals only)
USA
(Link Instruments,
Fairfield, NJ)
LINK Instruments
MSO-19
CHINA VIA
SINGAPORE
(Virtins
Technology)
VIRTINS
VT DSO-2820R
CHINA
(QingDao
Hantek
Electronics)
HANTEK
DSO-2250
BRAND,
MODEL
& COUNTRY
OF ORIGIN
pling rate only applies when a single
channel is being used. With both
channels in use, the figure drops to
100MSa/s – again, the sampling must
be interleaved. The effective analog
bandwidth would also drop to 40MHz.
The size of the 2820R’s internal
frame buffer memory is given as 40,000
bytes, which equates to 40,000 8-bit
samples using a single channel or
20,000 8-bit samples per channel when
using both channels. This is for normal real-time 8-bit sampling in frame
mode. However, Virtins claims that the
2820R and its other second generation
DSOs offer additional features, such as
hardware DSP-based bit-resolutionenhancement (BRE) and an adaptive
anti-aliasing filter.
The BRE feature is only available
for sampling rates below 100MSa/s
but when it is enabled the effective
sampling resolution of the 2820R increases by 1/2 a bit for each halving in
sampling frequency. So for a sampling
rate of 25MSa/s, BRE gives an effective
bit resolution of 9 bits, rising to 10 bits
at sampling rates below 6.25MSa/s, 11
bits at rates below 1.563MSa/s and so
on. In fact, the effective bit resolution
can be raised to a full 16 bits for sampling rates of 1.526kSa/s and below.
In other words, BRE provides a way
of trading sampling rate for effective
sampling bit resolution – allowing you
to examine and measure low-frequency signals with higher resolution.
Virtins MI software also allows for
equivalent time sampling and Streaming/Record mode sampling. The triggering flexibility is also quite good,
with Auto/Normal/Single shot or Slow
trigger modes, edge triggering (± slope),
adjustable triggering level and trigger
delay adjustable anywhere between the
start (-100%) and finish (+100%) of the
record length (normally the delay is
set to 0%, or the centre of the record).
You can also choose between either
of the two main input channels as the
trigger source when they’re both being
used. There’s also the ability to select
from 13 different trigger signal filtering
options including NIL (all pass), HFR
(high frequency reject), NR0-NR4 for
noise rejection, HN0-HN4 for HF rejection plus noise rejection, and HNX for
user-configurable filtering.
Voltage and time/frequency measurements (Vmax, Vmin, Vmean and
Vrms) are displayed automatically at
the top of the DSO display window,
while Vrms, frequency, duty cycle,
February 2015 73
x1 Probe Bandwidth
The blue curves in Fig.2 show the much poorer bandwidth of all three USBDSO devices when their input probes were switched into their x1 position. This
is quite typical – even though probe and scope manufacturers are usually rather
coy about this.
The reason for the reduced bandwidth of the probes in their x1 position is that
their frequency compensation is designed to optimise performance in the x10
position, where the input signal is divided by 10 before passing through the distributed capacitance, resistance and inductance of the output cable.
Since the input divider and compensation are effectively shorted out in the x1
position, the signals end up being attenuated by the output cable before they
reach the input of the scope.
So if you want to achieve the full bandwidth of any scope for measuring small
signals, it’s better to use a direct connection to the scope input – not a x10/x1
probe switched to its x1 position.
dBV, dBu, dB and dB(A/B/C filtering)
can be displayed in a separate DMM
window.
There doesn’t appear to be any provision for averaging, while the Chart
options seem to provide a choice of
Line, Scatter, Column, Bar and Step
so there is no option for sin(x)/x interpolation. However, you can select
a virtual display persistence of either
Zero, Phosphorescent, Rainbow or
Equivalent Time Sampling (with the
ability to set the number of frames).
You can also set the trace line width
from 1-10 pixels.
In addition, the DSO-2820R again
provides an FFT (Fast Fourier Transform) based Spectrum Analyser function, with nine selectable bin sizes
(128-32,768) and 50 different window
functions to suit many different applications. As well as Rectangle, Triangle, Hamming, Hanning, Blackman,
Exact Blackman, Blackman-Harris and
Blackman-Nuttal, Flat Top, Welch,
Riemann, Parzen and Bohman, there
are 14 different Kaiser windows, three
Poisson, three Hanning-Poisson, three
Cauchy, three Tukey, four Cosine and
three Gaussian windows.
Virtins quote the DSO-2820R’s maximum power consumption as 1.5W. We
measured the total current drain at
250mA when the DSO and FFT were
stopped and 322mA when they were
running. In other words, just over
half the drain of the DSO-2250 and
well within the capabilities of a single
standard USB 2.0 host port.
Link Instruments MSO-19.2
This is the smallest of the three
units and is housed in a moulded
74 Silicon Chip
plastic case measuring 92 x 61 x
20mm and weighing a mere 74g. Yet
Link Instruments has managed to
pack in a single-channel DSO with a
maximum sampling rate of 200MSa/s
and a claimed analog bandwidth of
60MHz, an 8-bit digital logic analyser
that can also sample at up to 200MSa/s
(synchronised with the DSO) with
decoders for SPI and I2C serial digital
signals, a 100MSa/s pattern generator or digital word generator, and an
FFT spectrum analyser covering from
DC to 100MHz. And it also has the
makings of a pulse-type time domain
reflectometer or TDR!
So as well as being very small and
light, it’s a versatile mixed signal package of PC-based test instruments. Small
wonder Link Instruments can boast
that the MSO-19 was chosen by NASA
as the only oscilloscope to be provided
on the International Space Station.
It comes with a single passive
switchable (10x/1x) 100MHz test probe
but there’s also a plug-in 15-wire harn
ess for the digital logic analyser inputs
and pattern generator outputs, together
with nine nanoclips for attaching the
digital leads to a device under test
(DUT).
There is also the USB cable, a 4-page
A5 installation guide, a 120mm CD
with the companion drivers and
software, plus a 100-page software
manual PDF.
Although the MSO-19.2 has a maximum real-time sampling rate of
200MSa/s, it also has provision for RIS
(Random Interleaved Sampling), a type
of Equivalent Time Sampling (ETS),
at rates of either 1GSa/s or 2GSa/s.
The interpolation seems to be fixed at
linear though, for both real time and
RIS sampling.
The size of the MSO-19.2’s internal
buffer memory appears to be 2KB,
with 1023 bytes allocated to the DSO
for storage of 1000 sample points and
the other 1023 bytes used for storing
the logic analyser and pattern generator data. When either of these buffers
become full, their contents are transferred to the PC via the USB cable.
There is a choice of Auto/Normal/
Single shot trigger modes, edge triggering (± slope), adjustable triggering
level over a range of ±4 full vertical
divisions with a resolution of 12.5mV,
and pulse width triggering. There’s
also an Autosetup mode.
Triggering can be from the DSO input channel or any of the eight logic
analyser digital inputs. The chosen
triggering signal also becomes available via the second BNC connector
on the front of the case, although this
connector can also be used as an output for the probe compensation signal,
a white noise signal, the TDR output
pulses or a low-frequency function
generator signal.
The measurements comprise Vmax,
Vmin, Vpp, Vmean, Vmedian and
Vrms, frequency, period, ± pulse
width and rise/fall times. In addition,
you can activate two voltage cursors
and two time cursors, to allow other
measurements.
The MSO-19.2’s FFT based spectrum analyser function seems to offer
a fixed number of bins but a choice of
seven window functions: Rectangular,
Tapered Rectangular, Triangular, Hamming, Hanning, Flat-Top or BlackmanHarris. There’s also a choice of display
types: Magnitude, Power Spectrum,
Power Density, Real or Imaginary.
There’s also the ability to produce and
display an average over 10, 20, 50 or
100 captures.
Total current drain of the sample
unit proved to be about 275mA when
the DSO and logic analyser were
stopped, and about 300mA when they
were running; well within the capabilities of a PC’s USB2.0 port.
Bench tests
Our first tests on each of the devices
were to check out their actual analog
bandwidths. We did this by installing
each one’s driver and software on a
Compaq CQ61 laptop running Windows 7 Home Premium (32-bit), and
then checking its HF response when
siliconchip.com.au
RELATIVE RESPONSE IN DECIBELS – DSO-2250
+5
+4
+3
RED CURVE:
Hantek DSO-2250 (x1 direct input, Rs = 50 Ω)
BLUE CURVE: Hantek DSO-2250 (x1 probe input, Rs = 50 Ω)
GREEN CURVE: Hantek DS0-2250 (x10 probe input, Rs = 50 Ω)
+2
+1
0
–1
–2
–3
–4
–5
100
kHz
200
300
400 500 600
800
1
MHz
2
3
4
5
6
7
FREQUENCY
8 9 10
MHz
20
30
40
50 60 70 80
20
30
40
50 60 70 80
20
30
40
50 60 70 80
100
MHz
RELATIVE RESPONSE IN DECIBELS – DSO-2820R
+5
+4
+3
RED CURVE:
Virtins DSO-2820R (x1 direct input, Rs = 50 Ω)
BLUE CURVE:
Virtins DSO-2820R (x1 probe input, Rs = 50 Ω)
GREEN CURVE: Virtins DSO-2820R (x10 probe input, Rs = 50 Ω)
+2
+1
0
–1
–2
–3
–4
–5
100
kHz
200
300
400 500 600
800
1
MHz
2
3
4
5
6
7
FREQUENCY
8 9 10
MHz
100
MHz
RELATIVE RESPONSE IN DECIBELS – MSO-19.2
+5
+4
+3
RED CURVE:
Link Inst’s MSO-19.2 (x1 direct input, Rs = 50 Ω)
BLUE CURVE:
Link Inst’s MSO-19.2 (x1 probe input, Rs = 50 Ω)
GREEN CURVE: Link Inst’s MSO-19.2 (x10 probe input, Rs = 50 Ω)
+2
+1
0
–1
–2
–3
–4
–5
100
kHz
200
300
400 500 600
800
1
MHz
2
3
4
5
FREQUENCY
6
7
8 9 10
MHz
100
MHz
Fig.2: these three graphs show the response of each instrument when measuring the unmodulated output of a Gratten
GA1484B signal generator. The generator’s output was set to +13dBm (1.0V RMS) and its signal fed to the input of the
device being tested via a 50Ω cable, terminated with a 50Ω load. Three different tests were done: (1) with the generator
signal fed directly into the DSO channel input; (2) with the signal fed via the matching test probe, set to the x10 position
and with its compensation optimised; and (3) with the signal via the same test probe set to the x1 position.
siliconchip.com.au
February 2015 75
Fig.3: a grab taken when the Virtins 2820R was being used to examine a 48MHz
0dBm sinewave using Equivalent Time Sampling. The Oscilloscope Properties
dialog is open at upper right, with the waveform visible to its left and the FFT
plot below.
Fig.4: this screen grab shows the Link MSO-19.2 capturing a 48MHz 0dBm
sinewave using RIS/ETS sampling at 1GSa/s. Note the linear interpolation and
the frequency measurement error.
Fig.5: the Link MSO-19.2 capturing a 20MHz +13dBm sinewave at 200MSa/s.
The linear interpolation is again quite evident.
measuring the unmodulated output of
a Gratten GA1484B signal generator.
The generator’s output was set to
76 Silicon Chip
+13dBm (= 1.0Vrms) and the output
signal taken to the input of the device
being tested via a 50Ω cable, terminated
with a 50Ω load at the device end to
minimise reflections and standing
waves. Three different tests were
done: (1) with the generator signal fed
directly into the DSO channel input; (2)
then via the matching test probe, set to
the x10 position and with its compensation optimised; and (3) finally, via
the same test probe to the x1 position.
The results of these tests are shown
graphically in Fig.2. In each case, the
curve for the first (direct) input test is
shown in red, that for the compensated
x10 test probe test is shown in green
and that for the x1 test probe test is
shown in dark blue.
You’ll see that the basic analog
bandwidth of the Virtins DSO-2820R
is within +0/-1dB up to 40MHz and
falls to -3dB at very close to 70MHz;
Similarly, the Link MSO-19.2 is within
+0.5/-0.5dB up to 44MHz and falls to
-3dB at about 74MHz. In this respect,
they’re both noticeably better than the
Hantek DSO-2250 which falls to the
-1dB point at about 8MHz and falls
to -3dB at close to 40MHz. Then after
falling to -4dB at 50MHz, it rises again
to reach +1.9dB at 100MHz.
We can probably attribute that upward rise in the DSO-2250’s red curve
to aliasing artefacts, so this part of the
curve is best ignored.
If we look at the green compensated
x10 probe curves, the results are a little
more equivocal. Even the response of
the Hantek DSO-2250 doesn’t fall to
-1dB until 24MHz and extends out
to 40MHz before it drops to -3dB. In
other words, the compensated x10
probe actually lifts the performance
of the DSO-2250.
On the other hand, the Virtins DSO2820R is now within +0.7/-1.0dB only
up to 34MHz, and falls to -3dB at about
38MHz – so the compensated x10
probe has lowered its performance a
little.
The Link MSO-19.2’s compensated
x10 probe has again improved its performance slightly, albeit with a small
dip between 40-50MHz and a small
peak at 60MHz. But its -3dB point has
risen to just on 80MHz.
The three blue curves show the x1
probe responses. These all show a significant drop in bandwidth compared
with the direct input bandwidth of
each device. This is to be expected
as it’s a fairly well known limitation
of the x1 position of just about all
switched passive probes.
But the really surprising thing when
siliconchip.com.au
you compare the three blue curves
in Fig.2 is that the bandwidth of the
MSO-19.2 with its probe in the x1
position is much better than the other
two. It rises to a broad +1dB peak at
15MHz and only falls to -3dB at about
28MHz. This improved performance
of the MSO-19.2’s probe in the x1
position suggests either that it’s of a
higher quality or that Link has built
some compensation into its software.
RIS/ETS sampling
As already noted, both the DSO2820R and the MSO-19.2 offer the
ability to perform RIS/ETS sampling
as well as real-time sampling, to allow
better definition of higher-frequency
repetitive signals. We tried out the
ETS functions of both devices but we
were not overly impressed with either
of them.
On the DSO-2820R, we initially
had a problem even getting the MI3.3
software to allow us to turn on the
ETS option in the Oscilloscope Chart
Options dialog – it seemed to be permanently greyed out and unavailable.
It was only after sending a help email
to Virtins that we received a reply
explaining that the trick was to set the
Trigger mode to Normal, the Trigger
Frequency Rejection to NIL and the
Trigger Delay to zero or less than zero.
We were advised that this is indicated on page 105 of the MI3.3 software manual and when we looked on
that page there it was – not in the text
though, just buried in a screen grab.
When we did get the ETS function
working, we were able to capture a
few waveforms that appeared to be
sampled at a higher sampling rate.
However there was no indication on
the screen of the effective sampling
rate – just the actual real-time sampling rate in the usual position near
the top of the screen.
In any case, the waveform being
displayed seemed to be infected with
travelling glitches, like those visible in
the screen grab of Fig.3 (which shows a
sinewave at 48MHz). This can also be
seen from the FFT display of the same
waveform below it where quite a few
spurs were also visible – although most
of them remained below about -55dBV.
At first we wondered if these
glitches and spurs were due to speed
limitations in the processor of the
laptop being used, which only has an
Intel dual-core Celeron CPU running
at 1.8GHz. So we tried installing the
siliconchip.com.au
Fig.6: here’s a grab of the Hantek DSO-2250 capturing a 35MHz +10dBm sinewave signal, with the acquisition menu showing that sin(x)/x interpolation has
been selected. Note the smoother sine waveform (green), plus the superimposed
FFT in deep purple.
MI3.3 software and Virtins driver on a
somewhat faster Asus BP6320 desktop
with an Intel Core i7-3770 CPU running at 3.4GHz, with Windows 7 Pro
(64-bit) and a 250GB SSD.
The results were almost identical,
so the glitches and spurs must be due
to something else.
When we tried out the RIS/ETS
function on the MSO-19.2, the results
were a little more promising. The
MSO-19.2’s software does show you
the effective sampling rate in this
mode, just below the horizontal speed
knob at upper left on the screen. And
there were no glitches as such on the
waveform display – just moving linear
interpolation vector lines and their
junctions, as you can see in the screen
grab of Fig.4.
As the displays using real-time sampling are much the same (see Fig.5), we
are inclined to think that the effect is
due to the lack of sin(x)/x sample interpolation in the MSO-19.2’s display
processing. Virtins MI3.3 and MI3.4
don’t seem to offer sin(x)/x interpolation either, so that may be part of the
explanation for those glitches.
to a rather dark purple which is hard
to read (see Fig.6).
There’s also a weird problem in the
FFT settings dialog box, where the
first and second harmonics are always
the same in both frequency and value
(Fig.7).
With the Link MSO-19.2 software,
there seems to be only a fixed number of FFT bins (“1000”, presumably
1024) but a choice of any of seven
window functions: Rectangular (Dirichlet), Tapered Rectangle (Tukey),
Triangular (Bartlett), Hamming, Hanning, Flat-Top and Blackman-Harris.
The spectrum can also be scaled in
Magnitude (mV), in Power Spectrum
(dBm), in Power Density or in its Real
and Imaginary components. Multiple
FFTs can also be averaged to give a
clearer spectrum display – see Fig.8.
The Virtins DSO-2820R + MI soft-
FFT/spectrum analyser functions
When we checked the FFT/spectrum analyser functions on each of the
three DSOs, there were strengths and
weaknesses in each one.
For example, Hantek’s DSO-2250
offers five bin sizes (from 256-4096)
and four window functions, plus trace
averaging and variable persistence.
Yet the spectrum display can only be
shown superimposed on the waveform
display and seems to be fixed in colour
Fig.7: this is the FFT Setting dialog
box in Hantek’s software for the
DSO-2250, showing how the window
function and number of bins can be
selected. Note the curious duplication
of the first and second harmonics
information.
February 2015 77
Nyquist-Shannon Sampling Theorem
You’ll find the name “Nyquist” cropping up frequently when you’re reading
about digital sampling, DSOs, ADCs and DACs – either by itself or together with
the name “Shannon”. That’s because Harry Nyquist and Claude Shannon were
two of the main researchers and theorists working in the field of information and
sampling theory early last century.
The sampling theorem usually named after them essentially defines the maximum signal frequency Fmax that can be conveyed by a digital sampling system
working at a sampling frequency Fs:
Fmax < Fs/2
By the way Fs is often called the Nyquist Rate, while Fs/2 is usually called the
Nyquist Frequency.
This looks simple enough but a couple of aspects need to be kept in mind when
you’re dealing with sampling. First, Fmax refers to the highest frequency COMPONENT in the signal being sampled. For example, a clean 99MHz sinewave signal
can be conveyed at a sampling rate of 200MSa/s and reconstructed faithfully at
the output – at least in theory. But the same can’t be done with a nominal 99MHz
signal having a more complex waveform, because this will have harmonics and
other components at frequencies well above 99MHz.
Remember – Fmax applies to ALL components in the signal, not just the
fundamental.
The other thing to remember is that the Nyquist-Shannon theorem assumes
that all samples are equally spaced in time. In other words, that Fs is fixed and
constant. This often doesn’t happen in the real world, where sampling clock signals
generally have at least a small amount of jitter.
The Nyquist-Shannon theorem tells you the effective analog bandwidth of a digital
scope by defining Fmax as below Fs/2. So with a DSO sampling at 200MSa/s,
Fmax will be just below 100MHz. But remember that this limit is (a) theoretical
and assumes no clock jitter and (b) applies to all frequency components in the
signal to be measured.
Digital scope makers often play safe by quoting a figure of Fs/2.5 for the effective
analog bandwidth of their instrument. But even this figure is really only relevant for
practical measurement of sinewave signals. When you want to examine square
waves or other signals with a high harmonic content, it’s best to assume that the
effective bandwidth is nearer Fs/10.
For example, you really need a digital scope sampling at 1GSa/s to examine
complex signals at frequencies up to 100MHz. And conversely, a USB scope
sampling at 200MSa/s has an effective analog bandwidth of around 20MHz for
complex signals.
ware doesn’t have any of the problems
of the DSO-2250 mentioned above. It
also has a larger choice of nine bin
sizes (from 128 - 32,768), plus those
50 different window functions. The
DSO and spectrum displays can also
be shown either separately on screen
or both together (one above the other).
But there does seem to be a lot of aliasing spurs on the spectrum display, as
shown in Fig.9.
Extra functions
In addition to its DSO and FFT/
Signal Analyser functions, the DSO2820R provides a signal generator
function of sorts. This makes use of the
same internal circuit used to generate
78 Silicon Chip
the 1kHz square-wave probe compensation adjust signal, available via the
small terminal lug on the DSO-2820R’s
front panel.
The signal generator is limited
to providing a digital or single bit
output – essentially either a square
wave or a maximal length sequence
(MLS), and with a fixed amplitude of
roughly 3.3Vp-p. However, it can be
programmed in terms of frequency,
via the DAC Device Setting dialog (accessed via the Setting -> DAC Device
menu options).
This dialog allows you to set the
generator to any frequency defined
by the expression 25MHz/N, where N
is any integer between 1 and 25,000.
You don’t have to work the divisor
out for yourself though; you simply
type in the frequency you want and
the MI software gives you the closest
frequency it can produce.
Clearly it can’t provide any frequency below 1kHz, because this is the
lowest frequency available (and just
happens to be the default frequency
used for adjusting probe compensation). But it is capable of providing
25,000 discrete frequencies, with
good resolution down at the low end
but gradually becoming poorer as you
move up.
At 10kHz the resolution is only
about 4Hz for example, while at
100kHz it has risen to about 400Hz.
Still, this could be useful in some applications.
The Link MSO-19.2 also provides a
number of extras, including the 8-bit
logic analyser with 100MHz bandwidth, SPI and I2C decoders, an 8-bit
pattern generator and a pulse-type
TDR. The latter again seems to be based
on the internal probe compensation
pulse generator, as its output emerges
from the same BNC output connector.
So all you need for pulse-type TDR
measurements is a short BNC-BNC
cable, a BNC T-adaptor and possibly a
coaxial series adaptor or two (for when
you want to check cables fitted with
connectors other than BNC).
Note that when used as a TDR, the
MSO-19.2 by itself can only be used
to check 50Ω cables. On the plus
side, it can convert delay times into
distances along the cable providing
you select the cable VOP (velocity of
propagation).
TDR comparisons
While we’re on the subject of TDR,
we did try out each of the three scopes
with our own Step-type TDR Dongle
as described in the December 2014 issue of SILICON CHIP. The results were
interesting.
The DSO2250 could display the
Step-type TDR waveforms quite nicely
but proved to be somewhat counterintuitive when it came to using its
vertical cursors to measure the delay
time between the start of the step and
any reflection of interest. Since there
is no real user manual and only a very
sketchy online help file, we had to
work out how to do it for ourselves.
But once we had done so it did give
quite useful results.
With the Virtins DSO-2820R, there
siliconchip.com.au
Useful Links
www.hantek.com/en/
www.hantek.in/en/
http://geoffg.net
www.virtins.com
www.multi-instrument.com
www.linkinstruments.com
http://shop1.usbdso.com
is no facility for moving the trigger
point in from the lefthand side of the
display (ie, no pre-trigger display).
This makes it difficult to be sure that
you are measuring from the start of
the TDR step. That aside, it proved
reasonably easy to make most TDR
timing measurements.
Although the MSO-19.2 does have
its own pulse TDR generator built in,
we found that it too would work with
our Step TDR Dongle. Not only that,
it would still convert reflection delay
times into distances along the cable
– providing you set the software into
its TDR mode and select the appropriate VOP.
However, the MSO-19.2 input has
a maximum full-scale vertical range
of 4V (±2V), so it can’t display the
full step waveform output of the TDR
Dongle when there’s either no cable
connected or the cable has an opencircuit somewhere. The trace simply
flies up to the top of the display and
stays there.
Fig.8: the FFT display when the Link MSO-19.2 was being used to examine a
48MHz 0dBm sinewave signal with sampling at 200MSa/s (real time sampling).
Fig.9: the FFT spectrum displayed by the Virtins 2820R (with the MI3.4 software)
when checking a 20MHz +13dBm sinewave signal. Note the relatively high harmonic peaks at 40, 60 and 80MHz (possibly due to front-end overload) and the
spurs at 10MHz, 30MHz and so on – probably caused by aliasing.
And the winner is?
That’s not easy to answer because
all three devices have their strengths
and weaknesses. If you mainly want
a 2-channel scope with the highest
possible bandwidth, the Virtins DSO2820R would probably be the winner.
If you want the highest possible
bandwidth but only need a single
scope channel, the Link MSO-19.2
would be your best bet.
Things get a bit more confusing if
you’re really looking for the most versatile FFT/Spectrum Analyser function.
Here you’d probably want to go with
the Virtins DSO-2820R and its MI software with nine bin sizes and choice of
50 window functions. But the Hantek
and Link devices and their software are
really not all that far behind when it
comes to many practical applications.
Finally, if you only need a single
siliconchip.com.au
Fig.10: here the MSO-19.2 was being used with our Step-TDR Dongle, to examine
an 18m-long cable terminated in 25 ohms. In its TDR mode, the MSO-19.2 can
even work out the cable distance corresponding to a reflection delay time.
channel scope but would also like the
added features of an 8-bit 200MSa/s
logic analyser, a 1023x8-bit 100MSa/s
pattern generator and a pulse-type
TDR, then go for the Link Instruments
SC
MSO-19.2.