This is only a preview of the February 2016 issue of Silicon Chip. You can view 39 of the 96 pages in the full issue, including the advertisments. For full access, purchase the issue for $10.00 or subscribe for access to the latest issues. Items relevant to "Micromite LCD BackPack With Touch-Screen Display":
Items relevant to "Solar MPPT Charger & Lighting Controller, Pt.1":
Items relevant to "Raspberry Pi Temperature/Humidity/Pressure Monitor, Pt.2":
Items relevant to "Valve Stereo Preamplifier For HiFi Systems, Pt.2":
Purchase a printed copy of this issue for $10.00. |
Defibrillators
Save Lives
CPR has long been considered the most
important first-aid skill you could learn.
Arguably it still is – but using an Automated
External Defibrillator (AED), is becoming just
as important. It can save many more lives!
Photo
courtesy
First Aid Plus
by Ross Tester
I
f we told you that performing CPR (Cardio-Pulmonary Resuscitation,
or heart-lung resuscitation) on someone who has suffered sudden
cardiac arrest, (SCA) is successful in only about 5-7% of cases, you’d
probably be surprised.
But you’d probably be even more surprised to learn that adding a defibrillator and CPR training shoots that success rate up to around 60%
(some authorities say even higher).
The reason is pretty simple: an SCA usually (though not always)
doesn’t just stop the heart pumping blood, often it “scrambles” the
electrical signals in the heart so that it does not rhythmically beat, even
if it wants to (and it does – the heart really wants to start rhythmically
beating again). An ECG of someone who has effectively died perhaps
half an hour before will sometimes show tiny pulses, albeit too small to
be of any use, as the heart valiantly tries to get going again).
What happens is that the confused electrical signals send the heart into
ventricular fibrillation, where it simply “quivers” rather than pumping
blood through its four chambers. Little or no vital oxygen-rich
blood gets to the heart and brain and within
a few minutes, the cells become
damaged. A few min-
14 Silicon Chip
siliconchip.com.au
Why publish this in SILICON CHIP?
If you’re wondering why we have published this feature in what is
primarily an electronics magazine, it’s for three very good reasons.
(1) Medical Electronics is a fascinating field that not too many
know about.
Most readers would have heard of defibrillators but we believe
that relatively few would know what they are actually used for (apart
from the “Packer Whacker” mentioned elsewhere), how they’re
used or their limitations. Indeed, most readers’ experience would
be the totally false movie and TV show images of a doctor looking
at a flat line on an ECG monitor, yelling “clear” and delivering a
shock which makes the patient almost jump off the bed!
As you read this story, you will begin to understand that defibrillators are highly sophisticated devices which these days do far
more than shock someone back to life after a heart attack (which,
by the way, they virtually never do unless that is followed by SCA
or sudden cardiac arrest!).
utes more and the damage becomes permanent and death
follows not long after.
The experts tell us that the first three to five minutes
are vital – that’s how long we have before damage starts
to occur. After ten minutes, assuming the victim isn’t
also suffering from hypothermia, or reduced temperature,
permanent damage is done. Hypothermia slows down the
damage but damage is still inevitable in time.
2: SILICON CHIP readership is overwhelmingy those who just might
need a defibrillator . . .
Whether they are in the position of first-on-the-scene after
someone has suffered a SCA or they are the person needing one,
we hope that this feature might open a few eyes to the appropriate
use of AEDs, now that they are to be found in many public buildings, offices and so on.
3: This is arguably the most important reason!
We hope that this article might encourage business owners and
managers to buy a defibrillator for their workplace (just as we have
now done at SILICON CHIP).
We list several defibrillators on page 19 – there are many more.
Today’s Public-Access AEDs are designed for totally novice and
untrained use – they tell you what to do and how to do it. Most are
relatively inexpensive these days (and tax deductible).
Just imagine how you would feel if someone in your office/factory/etc died because you hadn’t taken that simple step.
What’s wrong with CPR?
Absolutely nothing . . . everyone should learn it. CPR
is vital in the link to a successful resuscitation outcome,
providing oxygenated blood to the heart and brain. CPR
simply keeps the blood flowing where the heart isn’t functioning or not functioning properly.
The actions of the first aider repeatedly and rapidly
pushing down on the chest forces blood through the lungs
Doing it (almost!) for real: Manly Life Saving Club members Max Moon and Jonathan Curulli, under the watchful eye of
assessor Joe Mastrangelo, undertaking their AED qualification, part of Surf Life Saving’s Advanced Resuscitation Techniques
Certificate. The difference between this and the real thing is they’re using a “trainer” AED, obviously on a dummy!
siliconchip.com.au
February 2016 15
“Heart Attack” (MI; Myocardial Infarction) vs SCA (Sudden Cardiac Arrest)
A Sudden Cardiac Arrest (SCA), which can be defibrillated,
is quite different to a “heart attack”, more properly known as a
Myocardial Infarction (MI), which can’t.
Think of SCA as an “electrical problem” and MI as a “plumbing
problem”. SCA prevents the heart from beating due to its electrical signals being scrambled, while MI is something physical (for
example a blood clot or other blockage) in the arteries which
prevents blood flow to the heart itself.
The heart is a complex of nerves and muscles in their own right;
an MI will usually cause nerve and muscle damage when they are
starved of oxygen but will not necessarily be fatal if treated within
a reasonable time. Hence MI patients receive “bypasses” and
“stents” to allow blood to flow around or through a blocked area.
(to pick up oxygen) and to the heart and brain (especially),
along with the other organs of the body.
There are two problems with this: first of all, CPR is incredibly tiring for the person administering it; so much so
that the average person is lucky to effectively compress the
heart for two minutes. When you train in CPR, you train to
be able to swap with new people to keep going.
The rule for administering CPR is that it should be continued for as long as the first aider(s) is/are physically able
to continue, or until medical assistance arrives. Victims
have been reported to have recovered after more than an
hour of CPR – just so long as it is continued with no pauses.
Indeed, the record (for someone also suffering severe
hypothermia [low body temperature]) is a whopping six
hours, 30 minutes. Hypothermia appears to slow down the
onset of brain and other tissue damage.
(The opposite is also true, by the way – someone who is
overheated [hyperthermia] will tend to suffer organ damage
faster than someone who is cool.)
Good CPR is the only effective method of changing “fine
VF” (flatline) to “coarse VF”, which is a shockable rhythm.
However CPR will not stop VF – the quivering heart
needs an electrical shock to effectively stop the scrambling,
On the other hand an SCA if not treated immediately (<5 minutes)
will almost always be fatal .
An MI may in some cases may trigger an SCA but this is not usual.
As far as the first-aider is concerned, CPR, (preferably with
a defibrillator) is the only immediate means of treating an SCA,
whereas the usual treatment for MI is to sit the patient up and call
for medical help.
CPR must never be performed on a conscious (ie, MI) patient;
indeed modern defibs will not allow you to administer a shock if it
detects that a true pulse is present.
In all cases, MI or SCA, medical help should be summoned
immediately. If you are by yourself, do not stop CPR but yell out
until someone else is able to help
so that it can start it beating again in a “sinus” rhythm.
Enter the defibrillator
Believe it or not, the defibrillator’s job is to stop the heart.
It supplies a large shock current, directly across the heart,
to “freeze” the muscles and so stop it fibrillating. Then, the
heart may start beating properly of its own accord.
Even if it doesn’t, continuing CPR once the heart is
no longer fibrillating gives the victim the best chance of
survival.
Recent medical research suggests that the first shock is
the most important. Until now, most defibs “ramp up” the
shock in the belief that higher charges may damage the
heart. However, some manufacturers have now switched
over to delivering a large shock first up.
How much charge?
This depends a lot on the manufacturer and how their
waveform is set up. This, in turn, is determined by several
factors, not the least being the impedance between the pads
(ie, across the heart).
Early AEDs used a “monophasic” waveform – one where
the current went in one direction only. The recommended
charge for these was 360J for adult patients
(Joules = voltage x current x time). The
charge was delivered in a very short time
– a couple of milliseconds or so – meaning
that the other components, voltage and cur-
A defibrillator “Trainer”, such as being
used in the photo overleaf, looks and
works like a “proper” defib – except for
the vital detail that it cannot supply the
shock. The trainer gives the same prompts
as a real AED, including the spoken
instructions. It is clearly labelled as a
training device and the pads (also labelled
as training electrodes) can be used over
and over, unlike the real thing. This
trainer is from Cardac Science.
At right is the real defibrillator – notice
the difference? Apart from not having
the “Trainer” and “Not for Human Use”
labels, there isn’t one you can see.
16 Silicon Chip
siliconchip.com.au
rent, were rather high. Most worked on a voltage between
about 500V and 2000V DC.
Further research showed that a “biphasic” waveform,
where the current travelled in both directions, was just as
effective but with simpler circuitry, smaller battery and
lower weight. These days, the vast majority of AEDs which
you come across will be biphasic.
The advantage of biphasic waveforms is that the current can
be lower than monophasic, resulting in less potential damage
to the heart. And with lower power, there are fewer burns and
lower battery use. Not only that, the first shock success rate of
a biphasic machine is claimed to increase from 60% to 90%.
A typical AED might deliver 200 (or more) joules, with
a current of perhaps 30A or so delivered over 10ms.
The AED also analyses the “dryness” of the skin. With
drowning victims, the first aider is taught to dry the skin
before applying the pads. This is to both help the pads
“stick” and increase the impedance, so the AED can operate more efficiently.
There is a shock hazard when the shock is delivered,
so the first aider is also taught to ensure that no-one is in
contact with the victim (his/herself included!).
Note that manufacturers arrange delivery of their charges
differently so comparing one with another is not practical
nor accurate – none has been demonstrated to be superior
to another.
How do they work?
Public-access defibrillators (ie, those mounted in public areas and designed to be available to anyone, even
untrained) all give specific, clear instructions, including
where and how to place the pads. Apart from infants,
the location is almost invariably on the upper right chest
and the lower left side – this gives the best possible path
through the heart.
Infants usually have the pads placed front and back
over the heart.
Pacemakers and implanted defibs?
While it is quite possible that an external defibrillator will
“fry” the electronics in an implanted pacemaker or
defibrillator, the
Mr Bean’s Defibrillator
“Mr Bean Rides Again”
Tiger Aspect Productions
(1992)
You must have seen that episode of “Mr Bean” where Rowan
Atkinson attempts to deliver resuscitation to a man in the street
using a pair of jumper leads connected to a power pole.
Is it art imitates life or life imitates art? We’re not sure, but this
has some (OK, miniscule!) factual historical basis: early defibrillators (in hospital operating theatres) actually used 300-1000V
AC derived from a step-up transformer connected to the mains,
with research funded by the Edison Power Company!
The first successful use on a human (a 14-year-old boy being
operated on for a congenital chest defect) was in 1947 by Claude
Beck, professor of surgery at Case Western Reserve University
in Cleveland, Ohio. The boy’s chest was surgically opened and
manual cardiac massage was undertaken for 45 minutes before the
defibrillator arrived. Beck used paddles on either side of the heart.
Closed-chest defibrillation, using 100-150ms shocks <at> >1000V
AC, was pioneered in the USSR in the mid-1950s, while portable
defibrillators were first used in the late 1950s and early 1960s.
Today’s defibrillators deliver a very much shorter shock.
By the way, you can see the episode of “Mr Bean Rides Again”
on https://www.youtube.com/watch?v=OEwXQE5kh2SE
experts all say to ignore them.
If the victim is in ventricular fibrillation (VF), it’s obvious that an implanted defib is not doing its job. And a
pacemaker can’t work in VF anyway . . . the alternative is
to let the victim die.
Here’s the procedure
If you’ve done a first aid course, you’ll remember the
mnenonic
DRS ABCD
You may also remember that this stands for Danger,
Response, Send for Help, Airway, Breathing, Circulation
and Defibrillation.
The very latest teaching is much simplified and reflects
The “Packer Whacker”
25 years ago, media tycoon Kerry Packer suffered an SCA
while playing polo – and the one ambulance in NSW which
had a defibrillator on board happened to be the one which was
standing by for any injuries during the polo match.
Packer was revived using that defibrillator and he subsequently
donated enough money to the NSW Ambulance Service to equip
every ambulance with a defibrillator. These earned the nickname
“Packer Whackers” after their famous benefactor.
Packer died on Boxing Day, 2005, not from heart disease but
from kidney failure.
siliconchip.com.au
February 2016 17
the expert guidance that heart compressions are all-important to keep the
blood flowing; much more so than the
older routine, which experience has
shown that most people, even trained
first-aiders, can get wrong in the panic
of a “real” emergency.
But before anything else, as with
all first aid, you need to get someone
reliable to call for medical help (ie, an
ambulance).
The latest mnenonic, at least overseas (but could change here) is simply
CAB
This stands for Compressions, Airway and Breathing.
This calls for compressions to be
commenced as soon as no “signs of
life” are detected. These signs have
also been simplified – you no longer
have to feel for a pulse (that’s the main
thing lay people got wrong) but simply
establish that the victim is both unconscious and is not breathing.
Of course, you still need to ensure
the victim (and you!) are in no danger
and that the airway is clear but it is
now considered imperative to start
compressions as soon as possible – ie,
immediately!
While you’re doing this, have someone else set up the defibrillator.
The routine is:
1: Turn on the defibrillator – it will
go through a self-checking routine and
tell you when it’s ready to go (usually
only 5-10 seconds).
2: Attach the pads to the patient,
where shown by diagrams on the
defibrillator.
You will almost certainly need to
remove upper body clothing, wetsuits,
(most defibs contain a pair of scissors
to make this as quick as possible).
If the victim is a man with a very
hairy chest, use the shaver provided
to remove hair under the pad positions. If the victim is a female with
an underwire bra, this should also be
removed to eliminate any short circuit
possibility.
3: Plug in the pad leads to the defibrillator.
Two models of AED from the same company, the HeartOn A10 on the left and
the A15 on the right. The main difference is that the A15 caters for both adult
and child defibrillation without changing pads (courtesy APL Healthcare).
4: Follow the prompts that the defibrillator gives you (almost all these
days are spoken words).
It will tell you to do several things:
briefly stop CPR while it “analyses”
the electrical signals coming from the
heart via the pads.
It will probably tell you to continue
CPR, at the rate of 30 compressions
to two breaths, until it gives the next
instruction.
If it wants to deliver a shock, it will
tell you. Fully automatic defibs say
something like “stand clear. Delivering shock in 3-2-1 (seconds)”. Manual
defibs will tell you to “Shock now –
press red button”.
Even if you don’t know how to
perform CPR, in some cases the AED
will tell you – either by illustrations or
by voice commands. Some advanced
models will even tell you if your
compressions aren’t deep enough or
too slow.
Make sure that no-one is touching
the victim during analysis or shock
– the latter for obvious reasons; the
former because the minute electrical
signals in some else’s body may in fact
interfere with what the AED is trying
to read in the victim.
Then what?
Simply follow the instructions the
AED gives you. It may be that it advises
that no shock is possible, or that pulse
has been restored, or a variety of other
scenarios. Hopefully, by this time the
ambulance has arrived and they will
take over and treatment from you.
Buying an AED
As we said earlier, we hope that
this article may encourage businesses
and companies to invest in their own
AED and train some of their people
to use it. Make sure everyone knows
where it is!
Or at the prices of these units these
days, why not get together with a few
neighbours and buy one for home –
especially if you have older people
either living with you or close by?
Remember, though, that sudden
cardiac arrest is no respector of age –
whether by accident, trauma, disease
or illness, it can strike at any age!
Acknowledgement:
Our thanks to Gary Beauchamp, of
First Aid Plus, Sydney (02 9905 0155),
for assistance in preparing this feature.
On the left is a “normal” beating heart waveform, showing the rhythmical compression which pumps blood. This is referred to as a “sinus” rhythm. Compare this to the uncontrolled and basically useless waveform of a heart in ventricular
fibrillation. If not stopped (and that’s the job of a defibrillator) the lack of oxygenated bloodflow will quite quickly start to
cause damage to the heart muscles and to the brain (and other organs). Untreated, death is usually the end result.
18 Silicon Chip
siliconchip.com.au
Which AED is right for YOU?
AED Model Price (Dec15)
Replacements
(if known) Pads Battery
Choosing an AED is not all that simple: the best advice we can give is to look for
one which has a low “consumables” cost as well as an acceptable initial price.
Available From
Phone
Web
First Aid Plus
or Defibtech
(02) 9905 0155
1300 853 563
www.firstaidplus.com.au
www.defibtech.com.au
Australian First Aid
or APL Healthcare
1300 975 889
1300 727 580
www.australianfirstaid.com.au
www.aplhealthcare.com.au
HeartOn A15
$2250 (not known)
$2250
(not known)
Australian First Aid
or APL Healthcare
1300 975 889
1300 727 580
www.australianfirstaid.com.au
www.aplhealthcare.com.au
Heartsine samaritan PAD 500P (not known) (not known)
$2850 (not known)
$2860 (not known)
Aero Healthcare
or APL Healthcare
or Recovery Defibrillators
1800 628 881
1300 727 580
0413 223 472
www.aerohealthcare.com
www.aplhealthcare.com.au
www.recoverydefibrillators.com.au
Heartsine samaritan PAD 360P (not known) (not known)
Aero Healthcare
1800 628 881
www.aerohealthcare.com
Recovery Defibrillators
0413 223 472
www.recoverydefibrillators.com.au
APL Healthcare
1300 727 580
www.aplhealthcare.com.au
Defibtech Lifeline
HeartOn A10
$3135 (not known)
(not known) (not known)
$1950 $109.95 $199.00
$1950
(not known)
Heartsine samaritan PAD 350P
$2150
(not known)
Heartsine samaritan PAD 300P
$2400 (not known)
Laerdal HeartStart First Aid
$2390 $108
$250
(not known) (not known)
(not known) (not known)
Laerdal
or Australian Defibrilators
or St John
1800 331 565
1300 333 427
1300 360 455
www.laerdal.com.au
www.aeds.com.au
www.stjohn.org.au
Laerdal HeartStart Frx
(not known) (not known)
(not known) (not known)
$3200 $106
$250
Laerdal
or Australian Defibrilators
or St John
1800 331 565
1300 333 427
1300 360 455
www.laerdal.com.au
www.aeds.com.au
www.stjohn.org.au
First Aid Plus
(02) 9905 0155
www.firstaidplus.com.au
LifePak CR Plus
$2595
(not known)
Mindray Beneheart
(not known) (not known)
Australian Defibrilators
1300 333 427
www.aeds.com.au
Powerheart G3
(not known) (not known)
$2950 $90
$210
$2750 (not known)
Cardiac Science
or Australian Defibrilators
or Recovery Defibrillators
(03) 9429 2666
1300 333 427
0413 223 472
www.cardiacscience.com.au
www.aeds.com.au
www.recoverydefibrillators.com.au
Powerheart G5
(not known) (not known)
(not known) (not known)
(not known) (not known)
$2750 (not known)
Cardiac Science
First Aid Plus
or Australian Defibrilators
or Recovery Defibrillators
(03) 9429 2666
(02) 9905 0155
1300 333 427
0413 223 472
www.cardiacscience.com.au
www.firstaidplus.com.au
www.aeds.com.au
www.recoverydefibrillators.com.au
Schiller Fred Easy
(not known) (not known)
$1800
$100
Schiller Australia Pty Ltd
or Recovery Defibrillators
(02) 4954 2442
0413 223 472
www.schiller.com.au
www.recoverydefibrillators.com.au
Schiller Fred Easyport
(not known) (not known)
$2900 $100
$130
Schiller Australia Pty Ltd
or Recovery Defibrillators
(02) 4954 2442
0413 223 472
www.schiller.com.au
www.recoverydefibrillators.com.au
Zoll AED Plus
(not known) (not known)
$2795
(not known)
(not known)
(not known)
(not known)
(not known)
Zoll Medical Australia
or Defib Shop
or Australian Defibrilators
or St John
1800 605 555
1300 729 575
1300 333 427
1300 360 455
www.zoll.com.au
www.defibshop.com.au
www.aeds.com.au
www.stjohn.org.au
From our research, any of these AEDs appear to be quite suitable for office/factory/building use where untrained users may need to operate them.
However, this is not an exhaustive list. These days, most AEDs operate in a similar way but like any electronic device, might go about it their own way!
Some, for example, monitor CPR and will tell you if the compression depth is insufficient. Others may record data from the heart which can be used later
by a medical professional to review the treatment given and if necessary, tailor ongoing care.
The suppliers listed may or may not have stock and may have price changes, especially if they were on special offer when we checked.
Any prices shown are ex supplier’s websites. Many do not list prices of either the AEDs or their consumables (they want you to call them!).
SC
We suggest calling the numbers shown and/or visit their websites to determine features and availability.
siliconchip.com.au
February 2016 19
|